Invacare IOC100P user manual Operating Instructions

Page 14

SECTION 4—OPERATING INSTRUCTIONS

SECTION 4—OPERATING

INSTRUCTIONS

DANGER

NEVER smoke in an area where oxygen is being administered.

NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.

CAUTION

DO NOT use pediatric, low flow nasal cannulas or oxygen masks with the oxygen conserving regulator.

DO NOT block the cannula connection or kink cannula tubing when the oxygen conserving regulator is in use. This may damage the oxygen conserving regulator.

1.Prior to each use, inspect product for visible damage. DO NOT use if any damage is found.

WARNING

DO NOT direct flow of oxygen at any person, or flammable material when adjusting the flow dial.

2.Turn the selector switch on the oxygen conserving regulator to the “0” OFF position before opening the cylinder valve. Refer to FIGURE 4.2 on page 17.

3.To reduce the risk of rapid oxygen recompression and fire, open the cylinder valve slowly and completely so the pressure gauge moves slowly as it indicates the cylinder pressure.

Invacare® Element™

14

Part No 1160840

Image 14
Contents Invacare Element Pneumatic Conserver Accessories Warning Table of Contents Special Notes WordPrinciples of Operation Introduction General Guidelines Operating InformationGeneral Guidelines General Guidelines Features Model IOC100PConnection view of the IOC100P with CGA870 Installing the Conserving Regulator Cylinder post valve Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Typical Product Parameters RegulatorOxygen Cylinder Duration Maintenance Troubleshooting SolutionLimited Warranty USA