Invacare IOC100P user manual Typical Product Parameters, Regulator

Page 19

SECTION 6—TYPICAL PRODUCT PARAMETERS

SECTION 6—TYPICAL PRODUCT PARAMETERS

OXYGEN DELIVERY:

SELECTOR

LITER

 

SWITCH

FLOW

 

POSITION

EQUIVALENCY

 

 

 

 

0

OFF

 

 

 

 

1

1

 

 

 

 

2

2

 

 

 

 

3

3

 

 

 

 

4

4

 

 

 

 

5

5

 

 

 

 

6

6

 

 

 

CONTINUOUS FLOW

 

 

EMERGENCY BYPASS SETTING:

Factory Preset at 2 ± .5 L/min.

 

 

REQUIRED OPERATING

200 psi to 3000 psi

PRESSURE:

(13.8 bar to 206.8 bar)

 

 

REGULATOR:

Built-in, 25 ± 5 psi (1.7 bar ± .34 bar),

 

brass high pressure with aluminum low

 

pressure materials

 

 

 

DIMENSIONS:

 

 

 

 

Weight

Approximately 14.8 ounces (420 grams)

 

 

Overall Length

Approximately

 

5” L (12.7 cm) x 2.5” H (6.4 cm) x

 

2.5” W (6.4 cm)

 

 

 

OPERATING CONDITIONS

 

 

TEMPERATURE:

32°F (0°C) to 122°F (50°C)

 

 

 

OPERATING RELATIVE

 

 

HUMIDITY:

15% to 95%

 

 

OPERATING ALTITUDE:

0 to 10,000 feet (0 to 3,048 meters)

 

 

STORAGE/TRANSPORTATION:

Maximum -40°F (-40°C), 1% RH

 

Maximum 145°F (63°C), 44% RH

 

 

 

 

 

 

Part No 1160840

19

Invacare® Element™

Image 19
Contents Invacare Element Pneumatic Conserver Accessories Warning Table of Contents Word Special NotesPrinciples of Operation Introduction Operating Information General GuidelinesGeneral Guidelines General Guidelines Model IOC100P FeaturesConnection view of the IOC100P with CGA870 Installing the Conserving Regulator Cylinder post valve Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Regulator Typical Product ParametersOxygen Cylinder Duration Maintenance Solution TroubleshootingLimited Warranty USA