SECTION 5—REMOVING THE CONSERVING REGULATOR

SECTION 5—REMOVING THE CONSERVING REGULATOR

DANGER

NEVER attempt to remove the oxygen conserving regulator from an oxygen cylinder unless the cylinder valve is completely closed.

CAUTION

DO NOT try to loosen the T‐handle beyond the stop. Otherwise, damage to the oxygen conserving regulator will occur.

NOTE: For information on how to operate the cylinder, consult your equipment provider.

1.Completely close the oxygen cylinder valve.

2.Turn the oxygen conserving regulator selector switch to 2 L/min. continuous flow position.

3.Wait for oxygen to stop flowing from the oxygen conserving regulator.

4.Turn the T‐handle to loosen the connection to the post valve.

5.Remove the oxygen conserving regulator from the post valve.

Invacare® Element™

18

Part No 1160840

Image 18
Contents Invacare Element Pneumatic Conserver Accessories Warning Table of Contents Special Notes WordPrinciples of Operation Introduction General Guidelines Operating InformationGeneral Guidelines General Guidelines Features Model IOC100PConnection view of the IOC100P with CGA870 Installing the Conserving Regulator Cylinder post valve Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Typical Product Parameters RegulatorOxygen Cylinder Duration Maintenance Troubleshooting SolutionLimited Warranty USA