Use, Maintenance, and Care of the Devices

Handling and Storage

Handling and Storage

Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use.

Never store connectors loose in a box, desk, or bench drawer. This is the most common cause of connector damage during storage.

Keep connectors clean.

Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove.

Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.

3-10

85036B/E

Image 26
Contents User’s and Service Guide Manufacturing Part Number Print Date JuneDocumentation Warranty Contents Replaceable Parts Standard DefinitionsGeneral Information Kit Contents Calibration Kit OverviewEquipment Required but Not Supplied Installation of the Calibration DefinitionsCalibration Definitions AdaptersIncoming Inspection Calibration Kit Contents Serial Number Record for 85036E Recording the Device Serial NumbersSerial Number Record for 85036B Agilent 85036B kits with serial prefix 3102A and below Calibration Kits Documented in this ManualCalibration Kit History Agilent 85036B kits with serial prefix 3103A and belowClarifying the Sex of a Connector Preventive MaintenanceSpecifications Environmental Requirements Temperature-What to Watch Out ForEnvironmental Requirements Mechanical Specifications Mechanical SpecificationsSupplemental Mechanical Characteristics Mechanical CharacteristicsCertification Electrical SpecificationsElectrical Specifications Use, Maintenance, and Care of the Devices ESD Protection Setup Electrostatic DischargeLook for Obvious Defects and Damage First What Causes Connector Wear?Visual Inspection Inspect the Mating Plane SurfacesUse Compressed Air or Nitrogen Cleaning ConnectorsInspect the Precision Slotless Connectors female Clean the Mating Plane Surfaces ReinspectConnections How to Make a ConnectionFinal Connection Using a Torque Wrench Preliminary Connection Torque Wrench Information Wrench PositionsUsing the Torque Wrench How to Separate a Connection Connecting the Two-Piece Female OpenHandling and Storage Performance Verification Introduction How Agilent Verifies the Devices in This KitLimited Recertification Where to Send a Kit for RecertificationRecertification How Often to RecertifyRecertification 85036B/E Troubleshooting Troubleshooting Process Troubleshooting FlowchartReturning a Kit or Device to Agilent Contacting AgilentReturning a Kit or Device to Agilent 85036B/E Replaceable Parts Replaceable Parts for the 85036B Calibration Kit AdaptersReplaceable Parts for the 85036B Calibration Kit Replaceable Parts for the 85036E Calibration Kit Miscellaneous Items Introduction 85036B/E Standard Definitions Standard Class Assignments Table A-1 Standard Class AssignmentsBlank Form Table A-2 Standard Class Assignments Blank FormSetting the System Impedance Nominal Standard DefinitionsTable A-3 Standard Definitions Standard b Offset FreqTable A-4 Standard Definitions Blank Form Index Index-2 85036B/E Calibration