Agilent Technologies 85036B manual Calibration Definitions, Equipment Required but Not Supplied

Page 7

General Information

Calibration Kit Overview

Adapters

Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance. The adapters utilize a dual-beaded connector structure to ensure stable, repeatable connections. The beads are designed to minimize return loss and are separated far enough so that interaction between the beads is minimized.

The adapters are designed so that their nominal electrical lengths are the same, which allows them to be used in calibration procedures for non-insertable devices.

Calibration Definitions

The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.

The calibration definitions can be:

resident within the analyzer

entered from the front panel

Installation of the Calibration Definitions

The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.

If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.

Equipment Required but Not Supplied

Gages, torque and open-end wrenches, ESD protective devices, and various connector cleaning supplies are not included in the calibration kit but are required to ensure successful operation of the calibration kit. Refer to Table 6-3 on page 6-5for ordering information

85036B/E

1-3

Image 7
Contents Manufacturing Part Number Print Date June User’s and Service GuideDocumentation Warranty Contents Standard Definitions Replaceable PartsGeneral Information Calibration Kit Overview Kit ContentsAdapters Installation of the Calibration DefinitionsCalibration Definitions Equipment Required but Not SuppliedIncoming Inspection Calibration Kit Contents Serial Number Record for 85036B Recording the Device Serial NumbersSerial Number Record for 85036E Agilent 85036B kits with serial prefix 3103A and below Calibration Kits Documented in this ManualCalibration Kit History Agilent 85036B kits with serial prefix 3102A and belowPreventive Maintenance Clarifying the Sex of a ConnectorSpecifications Environmental Requirements Temperature-What to Watch Out ForEnvironmental Requirements Mechanical Characteristics Mechanical SpecificationsSupplemental Mechanical Characteristics Mechanical SpecificationsElectrical Specifications Electrical SpecificationsCertification Use, Maintenance, and Care of the Devices Electrostatic Discharge ESD Protection SetupInspect the Mating Plane Surfaces What Causes Connector Wear?Visual Inspection Look for Obvious Defects and Damage FirstInspect the Precision Slotless Connectors female Cleaning ConnectorsUse Compressed Air or Nitrogen Reinspect Clean the Mating Plane SurfacesPreliminary Connection How to Make a ConnectionFinal Connection Using a Torque Wrench ConnectionsWrench Positions Torque Wrench InformationUsing the Torque Wrench Connecting the Two-Piece Female Open How to Separate a ConnectionHandling and Storage Performance Verification How Agilent Verifies the Devices in This Kit IntroductionHow Often to Recertify Where to Send a Kit for RecertificationRecertification Limited RecertificationRecertification 85036B/E Troubleshooting Troubleshooting Flowchart Troubleshooting ProcessContacting Agilent Returning a Kit or Device to AgilentReturning a Kit or Device to Agilent 85036B/E Replaceable Parts Adapters Replaceable Parts for the 85036B Calibration KitReplaceable Parts for the 85036B Calibration Kit Replaceable Parts for the 85036E Calibration Kit Miscellaneous Items Introduction 85036B/E Standard Definitions Table A-1 Standard Class Assignments Standard Class AssignmentsTable A-2 Standard Class Assignments Blank Form Blank FormNominal Standard Definitions Setting the System ImpedanceStandard b Offset Freq Table A-3 Standard DefinitionsTable A-4 Standard Definitions Blank Form Index Index-2 85036B/E Calibration