Agilent Technologies E, 85036B manual Calibration Kit Overview, Kit Contents

Page 6

General Information

Calibration Kit Overview

Calibration Kit Overview

The Agilent 85036B and 85036E type-N calibration kits are used to calibrate Agilent network analyzers up to 3 GHz for measurements of components with 75Ω type-N connectors.

Kit Contents

The 85036B calibration kit contains the following:

one male and one female 75Ω type-N open termination

one male and one female 75Ω type-N short termination

one male and one female 75Ω type-N load

three 75Ω type-N to 75Ω type-N adapters

Refer to Table 6-1and Figure 6-1for a complete list of kit contents and their associated part numbers.

The 85036E calibration kit contains the following:

one male combination open/short termination

one male 75Ω type-N load

Refer to Table 6-2and Figure 6-2for a complete list of kit contents and their associated part numbers.

Broadband Loads

The broadband loads are instrument-grade, 75Ω terminations that have been optimized for performance up to 3 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.

Opens and Shorts

The opens and shorts are built from parts that are machined to the current state-of the-art precision machining.

The short’s inner conductors have a one-piece construction, common with the shorting plane. This construction provides for extremely repeatable connections.

The female open has a separate-piece inner conductor that is made from a low-dielectric-constant plastic to minimize compensation values.

Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The lengths of these offsets are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.

1-2

85036B/E

Image 6
Contents User’s and Service Guide Manufacturing Part Number Print Date JuneDocumentation Warranty Contents Replaceable Parts Standard DefinitionsGeneral Information Kit Contents Calibration Kit OverviewEquipment Required but Not Supplied Installation of the Calibration DefinitionsCalibration Definitions AdaptersIncoming Inspection Calibration Kit Contents Recording the Device Serial Numbers Serial Number Record for 85036BSerial Number Record for 85036E Agilent 85036B kits with serial prefix 3102A and below Calibration Kits Documented in this ManualCalibration Kit History Agilent 85036B kits with serial prefix 3103A and belowClarifying the Sex of a Connector Preventive MaintenanceSpecifications Temperature-What to Watch Out For Environmental RequirementsEnvironmental Requirements Mechanical Specifications Mechanical SpecificationsSupplemental Mechanical Characteristics Mechanical CharacteristicsElectrical Specifications Electrical SpecificationsCertification Use, Maintenance, and Care of the Devices ESD Protection Setup Electrostatic DischargeLook for Obvious Defects and Damage First What Causes Connector Wear?Visual Inspection Inspect the Mating Plane SurfacesCleaning Connectors Inspect the Precision Slotless Connectors femaleUse Compressed Air or Nitrogen Clean the Mating Plane Surfaces ReinspectConnections How to Make a ConnectionFinal Connection Using a Torque Wrench Preliminary ConnectionTorque Wrench Information Wrench PositionsUsing the Torque Wrench How to Separate a Connection Connecting the Two-Piece Female OpenHandling and Storage Performance Verification Introduction How Agilent Verifies the Devices in This KitLimited Recertification Where to Send a Kit for RecertificationRecertification How Often to RecertifyRecertification 85036B/E Troubleshooting Troubleshooting Process Troubleshooting FlowchartReturning a Kit or Device to Agilent Contacting AgilentReturning a Kit or Device to Agilent 85036B/E Replaceable Parts Replaceable Parts for the 85036B Calibration Kit AdaptersReplaceable Parts for the 85036B Calibration Kit Replaceable Parts for the 85036E Calibration Kit Miscellaneous Items Introduction 85036B/E Standard Definitions Standard Class Assignments Table A-1 Standard Class AssignmentsBlank Form Table A-2 Standard Class Assignments Blank FormSetting the System Impedance Nominal Standard DefinitionsTable A-3 Standard Definitions Standard b Offset FreqTable A-4 Standard Definitions Blank Form Index Index-2 85036B/E Calibration