Function

AC Volts1,2 -- True RMS (50 Hz to 500 Hz)

DC Volts

Continuity

Ohms

Diode test

Capacitance4

AC Amps5 -- True RMS (50 Hz to 500 Hz) (Models 111 and 112)

Range

6000 mV3

6.000 V

60.00V

600.0V

6000 mV3

6.000 V

60.00V

600.0V

600Ω

600.0Ω

6.000 kΩ

60.00kΩ

600.0kΩ

6.000 MΩ

40.00MΩ

2.200 V

1000 nF

10.00μF

100.0μF

10000 μF

10.00A continuous or 20 A overload for 30 seconds maximum

Resolution

1mV 0.001 V 0.01 V 0.1 V

1mV 0.001 V 0.01 V 0.1 V

1Ω

0.1Ω

0.001 kΩ

0.01kΩ

0.1kΩ

0.001 MΩ

0.01MΩ

0.001 V

1nF 0.01 μF 0.1 μF

1μF

0.01 A

Accuracy ± ( [ % of Reading ] + [ Counts ] )

Model 110

Model 111

Model 112

1.0 % + 3

1.0 % + 3

1.0 % + 3

0.7 % + 2

 

 

0.7 % + 2

0.7 % + 2

 

 

 

Beeper guaranteed on < 20 Ω, guaranteed off > 250 Ω; detects opens or shorts of 250 μs or longer.

0.9 % + 2

0.9 % + 2

0.9 % + 2

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

1.5 % + 3

1.5 % + 3

1.5 % + 3

0.9 % + 2

 

 

1.9 % + 2

1.9 % + 2

1.9 % + 2

1.9 % + 2

1.9 % + 2

1.9 % + 2

1.9 % + 2

1.9 % + 2

1.9 % + 2

100 μF - 1000 μF: 1.9% + 2

 

> 1000 μF: 10% + 90 typical

 

NA

1.5 % + 3

1.5 % + 3

 

 

 

12

Image 16
Contents Models 110, 111 Limited Warranty and Limitation of Liability Table of Contents Symbols Read Before Using the Meter Warnings and PrecautionsModels 110, 111 Unsafe Voltage Contacting FlukeTest Lead Alert Switch Position TerminalsDescription Measurement FunctionTrue RMS Multimeters DisplayModels 110, 111 MIN MAX AVG Recording Mode Display Hold To avoid electric shock, when Display Hold isWhen you apply a different voltage Backlight Model 112 OnlyButton Power-Up Options Power-Up OptionsManual Ranging and Auto Ranging Models 110, 111 Making Basic Measurements Making Basic Measurements Testing Diodes Using the Bar Graph CleaningReplacing the Battery andWFuse Testing the Fuse ModelsW111Specifications Resolution Accuracy ± % of Reading + Counts Function Models 110, 111