Fluke 110 user manual Accuracy ± % of Reading + Counts Function

Page 17

True RMS Multimeters

Specifications

Accuracy ± ( [ % of Reading ] + [ Counts ] )

Function

DC Amps

(Models 111 and 112)

Range

6.000 A

10.00A continuous or 20 A overload for 30 seconds maximum

Resolution

0.001 A

0.01 A

Model 110

NA

Model 111

1.0 % + 3

Model 112

1.0 % + 3

Hz6 (V or A input )

99.99Hz

999.9Hz

9.999 kHz

50.00kHz

0.01Hz

0.1Hz

0.001 kHz

0.01kHz

0.1 % + 2

0.1 % + 2

0.1 % + 2

MIN MAX AVG Accuracy and Response Time

Accuracy is the specified accuracy of the measurement function ± 12 digits for changes >275 ms in duration (± 40 digits in AC). Typical response time: 100 ms to 80 % of signal, except V AC and A AC.

1.AC voltage ranges are specified from 5% of range to 100% of range.

2.Crest factor of 3 at full scale up to 300 V, decreasing linearly to crest factor 1.5 at 600 V.

3.The 6000 mV range can only be entered in Manual Range mode.

4.For film capacitors.

5.Crest factor of 3. AC current is not specified below 3A.

6.Hz is specified from 5 Hz to 50 kHz in volts, from 50 Hz to 5 kHz in amps.

Function

Volts AC

Volts DC

Ohms

Diode test

Input Impedance (Nominal)

>5 MΩ < 100 pF

>10 MΩ < 100 pF

Open Circuit Test Voltage

<1.5 V DC

2.4to 3.0 V DC

Common Mode Rejection Ratio

> 60 dB at DC, 50 Hz or 60 Hz

> 100 dB at DC, 50 Hz or 60 Hz

 

Full Scale Voltage

To 6 MΩ

 

40 MΩ

< 600 mV DC

 

< 1.5V DC

 

2.400 V DC

Normal Mode Rejection

> 50 dB at 50 Hz or 60 Hz

Short Circuit Current

<500 µA

1.2mA typical

13

Image 17
Contents Models 110, 111 Limited Warranty and Limitation of Liability Table of Contents Models 110, 111 Read Before Using the Meter Warnings and PrecautionsSymbols Test Lead Alert Contacting FlukeUnsafe Voltage Terminals Switch PositionDescription Measurement FunctionDisplay True RMS MultimetersTo avoid electric shock, when Display Hold is Models 110, 111 MIN MAX AVG Recording Mode Display HoldWhen you apply a different voltage Backlight Model 112 OnlyManual Ranging and Auto Ranging Power-Up OptionsButton Power-Up Options Models 110, 111 Making Basic Measurements Making Basic Measurements Testing Diodes Cleaning Using the Bar Graph Testing the Fuse ModelsW111 Replacing the Battery andWFuseSpecifications Resolution Accuracy ± % of Reading + Counts Function Models 110, 111