Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the unit.

Problem

Possible Cause

Solution

 

 

 

No display.

Power not connected or

Ensure the power cable is

 

switch at back is not on.

connected and that the

 

 

switch at the back of the

 

 

unit is in the ON position.

Insufficient air

Ramp Time is in use.

delivered from the

 

VPAP.

 

Wait for air pressure to build up.

 

Air filter is dirty.

Replace air filter.

 

Air tubing is kinked or

Straighten or replace

 

punctured.

tubing.

 

Air tubing not connected

Check air tubing.

 

properly.

 

 

Mask and headgear not

Adjust position of mask

 

positioned correctly.

and headgear.

 

Plug(s) missing from

Replace plug(s).

 

access port(s) on mask.

 

 

Pressure required for

See your clinician to

 

treatment may have

adjust the pressure.

 

changed.

 

 

 

 

The VPAP does not

Power cord not connected

Connect power cord

start when you

properly.

firmly at both ends.

breathe into the

 

 

mask.

Power outlet may be

Try another power outlet.

 

faulty.

 

 

The VPAP unit not

Switch power switch at

 

switched on.

rear of the VPAP to ON.

 

SmartStart not on.

Enable SmartStart.

 

Mask Alarm has been

Disable Mask Alarm to

 

enabled; SmartStart has

enable SmartStart.

 

automatically been

 

 

disabled.

 

Troubleshooting 31

Image 38
Contents 248127/1 05 Illustrations 52 cm 3 m English Manufactured by Contents Limited Warranty Index User/Owner Responsibility IntroductionMedical Information What the Vpap III and Vpap III ST are Intended forPage Adverse Effects Medical InformationPage Vpap System MasksAccessories HumidifiersAttaching a Humidifier Setting up the Vpap SystemSetting Up The Vpap Attaching a HumidAire Humidifier Attaching a HumidAire 2i HumidifierAttaching a HumidAire 2iC Humidifier Attaching a Passover Humidifier Keypad Keys Using the LCD Screen and KeypadLCD Screen KeyFunction FrontKey Function Left green Right redPage Starting Treatment Starting TreatmentRamp time Stopping Treatment Using the HumidAire 2i Warm-Up FeatureUsing the Mask-Fitting Feature High Leak Page Weekly Cleaning Cleaning and MaintenanceDaily Cleaning Periodic CleaningReplacing the Air Filter ServicingRamp Screen How to Use the Vpap MenusHow to Use the Standard Menu How to Use the Detailed Menu if enabled by your clinician To exit out of the menu Press the Right key exitVpap Detailed Menu Series if enabled by your clinician To enter a submenu Press the Left key enter Settings MenuMenu Functions Detailed menu only Function Default Function Description Setting Options Settings for mask types MaskResults Menu Settings MaskResults Menu Function Function Description Setting Options Options Menu Servicing MenuTreatment Screens Treatment screenHow to Use the Vpap Menus Page Starting out Using a Battery and an InverterHelpful Hints Nasal IrritationPage Troubleshooting TroubleshootingProblem Possible Cause Solution Check tube Key if doneLast session Page System Specifications Dynamic pressure characteristicsIEC 60601-1 Classifications Flow sensor in flow generatorPressure cm H2O Pressure Volume curvePressure Variation BreathsSymbols which appear on the product Environmental informationEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Limited WarrantyProduct Warranty Period Page Index