Guidance and manufacturer’s declaration – electromagnetic immunity (Continued)
The VPAP is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP should assure that it is used in such an environment.
Immunity | Compliance | Electromagnetic environment – | |
test | test level | level | guidance |
|
|
|
|
|
|
| Portable and mobile RF communications |
|
|
| equipment should be used no closer to any |
|
|
| part of the VPAP, including cables, than the |
|
|
| recommended separation distance |
|
|
| calculated from the equation applicable to |
|
|
| the frequency of the transmitter. |
|
|
| Recommended separation distance |
Conducted RF | 3 Vrms | 3 Vrms | d = 1.17 √ P |
IEC | 150 kHz to 80 MHz |
|
|
Radiated RF | 10 V/m | 10 V/m | d = 0.35 √ P 80 MHz to 800 MHz |
IEC | 80 MHz to 2.5 GHz |
|
|
|
|
| d = 0.70 √ P 800 MHz to 2.5 GHz |
|
|
| where P is the maximum output power |
|
|
| rating of the transmitter in watts (W) |
|
|
| according to the transmitter manufacturer |
|
|
| and d is the recommended separation |
|
|
| distance in meters (m). |
|
|
| Field strengths from fixed RF transmitters, as |
|
|
| determined by an electromagnetic site |
|
|
| survey,a should be less than the compliance |
|
|
| level in each frequency range.b |
|
|
| Interference may occur in the vicinity of |
|
|
| equipment marked with the following |
|
|
| symbol: |
|
|
|
|
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the VPAP is used exceeds the applicable RF compliance level above, the VPAP should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the VPAP.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.