3.1.1.3 Status LEDs
ml/min When lit, the digital display shows flow rate in mL/min.
PSI When lit, the digital display shows system pressure in psi.
HI PR When lit, the display shows the
LO PR When lit, the display shows the
RUN Lights to indicate that the pump is running.
FAULT Lights when a fault occurs and stops the pump.
3.1.1.2
Non-volatile Memory Reset: If the pump is operating erratically, there is the possibility that the memory has been corrupted. To reset the memory and restore the pump to it's default parameters, press and hold the UP-ARROW button when the power is switched on. Release the button when the display reads "rES". The parameters stored in non-volatile memory, i.e., the flowrate, the pressure compensation, the voltage/frequency select, the lower pressure limit, and the upper pressure limit will be set to the factory default values. The head type setting is the only parameter not changed by the non-volatile memory reset function. If the firmware is upgraded to a newer version, a non- volatile memory reset will automatically occur the first time the power is switched on.
3.1.1.5
Display Software Version Mode: The software version can be displayed during