SSI America 90-2581 REV B manual A.1.3, Command Interpreter, “/” character

Page 40
A.1.3

output is at a low logic level. A low logic level is -3.0 to -15 volts and a high logic level is 3.0 to 15 volts.

 

A.1.3

Command Interpreter

 

The pump’s high-level command interpreter receives and responds

 

to command packets. The pump will not send a message except when

 

prompted, and it will send a response to every valid command as

 

described below. The response to an invalid command is “Er/”.

 

Each command is characterized by a unique two-letter command

 

code, and only one command can be issued per line. Case is not

 

important; that is, the command codes “PR” “Pr” “pR” and “pr” are all

 

equivalent. Response strings sent by the pump are terminated by the

 

“/” character.

 

If the pump's response is "Er/", send a "#" to clear any characters

 

which may be remaining in the command buffer. The pump will

 

automatically clear all characters in the command buffer after one

 

second elapses from the time at which the last character of an

 

incomplete command was sent.

 

The command packets are as follows:

Command

Response

Comments

RU

OK/

Sets the pump to the RUN state.

ST

OK/

Sets the pump to the STOP state.

FLxxx

OK/

Sets the flow rate to x.xx or xx.x mL/min where the range

 

 

is fixed for the pump head size, i.e., for 0.01 to 9.99

 

 

mL/min xxx = 001 to 999, for 0.1 to 39.9 mL/min xxx =

 

 

001 to 399.

FOxxxx

OK/

FMxxxx

OK/

PR

OK,x/

 

(x, xx, xxx, or xxxx)

CCOK,x,y.yy/

(x, xx, xxx, or xxxx) (y.yyy, y.yy, yy.yy, or yy.y)

Sets the flow rate to xx.xx or xxx.x mL/min where the range is fixed for the pump head size, i.e., for 0.01 to

10.00mL/min xxxx = 0001 to 1000, for 0.1 to 40.0 mL/min xxxx = 0001 to 0400.

Sets the flowrate to x.xxx mL/min, i.e.,

for 0.001 to 9.999mL/min xxxx = 0001 to 9999. for 10.00 to 12.00mL/min xxxx = 1000 to 1200. Reads the pump's current pressure, where:

x, xx, xxx, or xxxx = Current pressure in PSI Reads the pump's current pressure and flowrate, where:

x, xx, xxx, or xxxx = Current pressure in PSI y.yyy, y.yy, yy.yy, or yy.y = Flow rate in mL/min

The format is y.yy and yy.yy for a standard pump head, y.yyy for micro pump head or yy.y for a macro pump head.

A-2

Image 40
Contents Binary Solvent Delivery Module Operators ManualPage 1.INTRODUCTION 1.1.1 Pump Features fFast & easy setup1.1.3 Self-FlushingPump Head 1.1.2 Wetted Pump MaterialsMicroprocessor advanced control 1.1.4 Self-Flushand Seal Life It is recommended that the Self Flush feature be used to improve seal life in a number of applications. In particular, as stated above if pumping Buffers, Acids/Bases or any inorganic solution near saturation, the pump should utilize the Self Flush feature. With every piston stroke, an extremely thin film of solution is pulled back past the seal. If this zone is dry without use of Self Flush, then crystals will form with continuous operation, which will ultimately damage the sealFlow Rates 2.INSTALLATION 2.1Unpacking and Inspection2.2Location/Environment 2.3Fluid Connections & PrimingSELF-FLUSH PUMPConnect wash solution inlet and outlet tubing opaque to the flush housing as shown2.4Electrical Connection 2.5.1 Solvent Out-gassingand Sparging2.5Solvent Preparation 2.6.1Mobile Phase Reservoirs 2.5.2 Cavitation2.5.3 Filtration 2.5.4 Initial system pressurization Daily2.6.2 Self-FlushSolution 2.6.3 Inlet Tubing and Filters2.6.4 Priming the Pump and the Flushing Lines 2.6.6 Long Term Pressure Calibration Accuracy2.7.2 Packaging for Shipping 2.7.1Isopropanol Flush2.7Preparation for Storage or Shipping 3.OPERATION 3.1.1Control PanelNon-volatileMemory Reset If the pump is operating erratically, there is the possibility that the memory has been corrupted. To reset the memory and restore the pump to its default parameters, press and hold the UP-ARROWbutton when the power is switched on. Release the button when the display reads rES. The parameters stored in non-volatilememory, i.e., the flowrate, the pressure compensation, the voltage/frequency select, the lower pressure limit, and the upper pressure limit will be set to the factory default values. The head type setting is the only parameter not changed by the non-volatilememory reset function. If the firmware is upgraded to a newer version, a non- volatile memory reset will automatically occur the first time the power is switched on 3.6Symbols 3.5Rear Panel Remote Input4.THEORY OF OPERATION 4.1.1Liquid System Flow Path4.1.2 Pump Cycle 4.1.3 Pulse Damping4.2.2 DC Power Supply 4.2.1 Microprocessor Control4.2.3 Remote Interfacing 4.2.4 Motor Stall DetectorPage 5.MAINTENANCE 5.1.1 Inlet Filters5.2.1Removing a Pump Head 5.1Filter Replacement3.Remove the inlet line from the inlet check valve OUTLET FLUSHING CHECK VALVE OUTLET FLUSHING 5.2.2 Cleaning the Pump Head Assembly 1.Inspect the piston seal cavity in the pump head. Remove any foreign material using a cotton swab, or equivalent, and avoid scratching the sealing surfaces. Repeat for the self-flushhousing. Be sure no fibers from the cleaning swab remain in the components3.If the check valves have been removed, tighten the check valves on stainless steel pumps to 75 inch-poundsor enough to seal at maximum pressure. Do not exceed maximum torque. For Bioclean PEEK pumps, tighten each check valve firmly by hand. Do not go ¼ turn past finger tight 5.2.3 Replacing Piston Seals Lower than normal pressure, pressure variations, and leaks in the pumping system can all indicate possible problems with the piston seal. Depending on the fluid or mobile phase used, piston seal replacement is often necessary after 1000 hours of running time1.Remove the pump head as described in Section 5.2.1. Use caution so as not to damage the sapphire piston 5.2.4 Changing the Piston 5.Condition the new seal as described in Section5.2.5Replacing the Pump Head 5.3Conditioning New Seals5.5.1Removing the Pulse Damper 5.5.2 Pulse Damper Refurbishing5.4Check Valve Cleaning and Replacement 5.5Pulse Damper Replacement5.5.3 Pulse Damper Installation 5.6Cleaning the Pump5.8Other Pump Maintenance 5.7Lubrication7-14 5.9Fuse Replacement 1.Unplug the unit 2.Remove the cover 6.PROBLEM SOLVING Quick Guide to Problem SolvingYou Notice This May Mean7.LIST OF REPLACEMENT PARTS 880201 Seal Kit, Aqueous, 5mL APPENDIX A A.1.1 Hardware ImplementationPart Description Part NumberA.1.3 Command Interpreter“/” character Page A.2.1 Pressure Fault and Motor Stall Fault Output A4-pinterminal board connector and a 10-pinterminal board connector are provided on the back panel. Any device capable of providing the proper run/stop logic level, flowrate control frequency, or flowrate control voltage can be used as a remote controlling device for pump operation via this connection. The terminal board connectors can be removed for ease of connecting wires, if desired, by pulling firmly rearward and should be reinserted firmly afterwardA.2.2 General Information on Inputs the EVENT 1 terminal and the EVENT 2 and EVENT 3 terminals is affected. EVENT 2 is Normally Closed connected to EVENT 1 until a fault occurs and then opens. EVENT 3 is Normally Open not connected to EVENT 1 until a fault occurs and then closesA.2.4 Run and Stop Inputs The PUMP-RUN, PUMP-STOP,and ENABLE IN inputs operate from an internal 5 VDC source and each one draws approximately 0.008 amps when connected to COM. To activate either the PUMP- RUN, PUMP-STOP,or ENABLE IN input connect it to COM. Any device capable of switching 0.008 amps can be connected between the PUMP-RUN, PUMP-STOP,or ENABLE IN input and COM, such as: a switch contact, a relay contact, an open collector output, an open drain output, or any output with a high logic level output of 3.8 to 6.0 volts and a low logic level output of 0.0 to 0.5 volts. A switch contact or a relay contact is preferred since this type of connection will provide isolation between the pump and the controlling device. The COM terminal is internally connected to the pumps chassis ground and should be connected to the controlling devices ground or zero volt terminal when the controlling device has an open collector output, an open drain output, or any output with logic level outputA.2.4 Enable Input When activated ENABLE IN is at a low logic level, the ENABLE IN input disables flowrate control on the front panel and enables flowrate control on the back panelAPPENDIX B EZ GRIP FITTING GUIDEScientific Systems Inc Warranty Statement A-10