APPENDIX A
A.1 Rear Panel Serial Communications Port - Pump
An
A.1.1 Hardware Implementation
The REMOTE INPUT serial communications port is configured for 9600 baud, 8 data bits, 1 stop bit, and no parity. The connector is a standard
Pin | Function |
1, 6 | Ground |
2 | DSR (Handshaking input to pump) |
3 | RXD (Serial data input to pump) |
4 | TXD (Serial data output from pump) |
5 | DTR (Handshaking output from pump) |
Special wiring considerations: Use the following chart for interfacing the pump's serial communications port to either a
Pump (RJ11) | Signal | IBM (DB25)a | IBM (DB9)b |
1, 6 | Ground | 7 | 5 |
2 | DSR | 20 | 4 |
3 | RXD | 2 | 3 |
4 | TXD | 3 | 2 |
5 | DTR | 6 | 6 |
a Jumper pins 4, 5, and 8 on DB25. b Jumper pins 1, 7, and 8 on DB9.
Part Description | Part Number |
Modular Cable | |
Adapter | |
Adapter |
A.1.2 Hand-Shaking
The pump uses hardware handshaking. The pump will not transmit on the TXD output if the DSR input is at a low logic level. And, the pump will not receive on the RXD input when the DTR