Honeywell MicroDock II manual Introduction, Safety Information Read First

Page 3

Introduction

aWarning

To ensure personal safety, read Safety Information - Read First before using the MicroDock II base station.

The MicroDock II Automatic Test and Calibration Station (“the base station”) provides automated calibration and bump testing for the GasAlert Extreme, GasAlertClip Extreme, GasAlertMi- cro, GasAlertMicro 5/PID/IR, GasAlertMicroClip, and GasAlert- Max XT detectors. The system is expandable to include up to 10 docking modules (maximum six charging docking modules plus four non-charging docking modules.

Safety Information - Read First

Use the station only as specified in this guide. Read the follow- ing Cautions before using the station.

aCautions

If the base station is damaged or parts are missing, contact BW Technologies by Honeywell immediately.

This equipment uses potentially harmful gas for calibration. Use in a well-ventilated area only.

MicroDock II

aCautions

The base station must be attached to a venting system or the base station must be used in a well-ventilated area.

Do not immerse the station in liquids.

The maximum recommended exhaust line length is

15.24m (50 ft.).

Ensure that the inlet filter is clean.

Ensure that all gas cylinders contain enough gas.

A demand flow regulator must be used with all gas cylinder connections.

Calibrate and bump test only in a safe area that is free of hazardous gas.

Do not expose the station to electrical shock or severe continuous mechanical shock.

The base station warranty will be void if the unit is disassembled, adjusted, or serviced by non-BW Technologies by Honeywell personnel.

Ensure the exhaust line is not connected to a negative pressure system.

1

Image 3
Contents Quick Reference Guide USA Canada Europe +440 1295 Contacting BW Technologies by HoneywellIntroduction Safety Information Read FirstPushbuttons Display ElementsDescription Display User Options MenuInstallation Installation Inserting the Detector Infrared or intense ambient light sun or halogen mayInterfere with the base station/detector communication Using the Base Station Using the Base StationAre not setup correctly Bump Test Press C exit to return to normal operationCalibration excluding GasAlertClip Extreme MicroDock Maximum Storage Capacity EventloggingTo charge the battery pack, complete the following Charging the Battery Pack OptionalMaintenance No user-serviceable parts insideRequired. To order replacement parts, refer to Specifications Humidity 0 to 50% Charger SpecificationsSafety Standards must be followed Charger system ingress protection IP20MicroDock