Honeywell MicroDock II manual Using the Base Station

Page 9

MicroDock II

Using the Base Station

When the detector has been inserted correctly, the RUN LEDs light yellow, Unit Inserted displays on the base station LCD, and MicroDock displays on the detector LCD. The base station LCD displays the docking mod- ule number, and the type and serial number of the detector.

GasAlertMicroClip

Important: If the GasAlertMicroClip is fitted with a calibration cap or an auxiliary filter, it must be removed prior to inserting it into the docking module. Refer to the GasAlertMicroClip User Manual and the MicroDock II Base Station User Manual.

Insert the bottom of the detector (serial number face up) at a 30° angle into the detector bay.

Note

The docking module lid only raises upward 30°. Do not force the lid beyond its limit.

Close the lid and press until the release tabs click.

When the detector has been inserted correctly, the RUN LEDs on the docking module light yellow and Unit Inserted displays on the base station LCD. The base station LCD displays the docking module number, and the type and serial number of the detector.

GasAlertMax XT

Insert the bottom of the detector (serial number face up) at a 30° angle into the detector bay.

Note

The docking module lid only raises upward 30°. Do not force the lid beyond its limit.

Close the lid and press until the release tabs click. Push the pump connector towards the detector pump.

Ensure green is visible in the lower indicator in the pump connector. If it is not visible, the MicroDock II base station will not recognize the detector.

Note

The detector will briefly alarm when the pump connector is indicated. This is normal. The alarm will deactivate when the detector is properly inserted into the docking module.

When the detector has been inserted correctly, the RUN LEDs on the docking module light yellow and Unit Inserted displays on the base station LCD. The base station LCD displays the docking module number, and the type and serial number of the detector.

Using the Base Station

aWarning

To prevent possible personal injury and/or property damage, only use the station in a safe area that is free of hazardous gas.

Ensure that the station is attached to a venting system or used in a well ventilated area.

7

Image 9
Contents Quick Reference Guide USA Canada Europe +440 1295 Contacting BW Technologies by HoneywellIntroduction Safety Information Read FirstPushbuttons Display ElementsDescription Display User Options MenuInstallation Installation Inserting the Detector Infrared or intense ambient light sun or halogen mayInterfere with the base station/detector communication Using the Base Station Using the Base StationAre not setup correctly Bump Test Press C exit to return to normal operationCalibration excluding GasAlertClip Extreme MicroDock Maximum Storage Capacity EventloggingTo charge the battery pack, complete the following Charging the Battery Pack OptionalMaintenance No user-serviceable parts insideRequired. To order replacement parts, refer to Specifications Safety Standards must be followed Charger SpecificationsCharger system ingress protection IP20 Humidity 0 to 50%MicroDock