Page 84

Sensor Data Register

Register 0x0016h

R

82 bytes

 

 

 

 

 

 

Variable

Type

Length

Description

 

 

 

 

 

 

Pressure

Float

4

Manifold Pressure is PSIA

 

 

 

 

P_Volts

Float

4

Pressure sensor output Voltage

 

 

 

 

Vacuum_P

Float

4

Pressure with all value closed and pump on in PSIA

 

 

 

 

Ambient_P

Float

4

Absolute Ambient Pressure in PSIA

 

 

 

 

Box_T

Float

4

Enclosure Temperature in Degrees C

 

 

 

 

Box_T_Volts

Float

4

Box temp sensor output voltage

 

 

 

 

Bench_T

Float

4

Optical bench temperature in Degrees C

 

 

 

 

Bench_T_Volts

Float

4

Bench temp sensor output voltage

 

 

 

 

Bench_Z_T

Float

4

Optical bench temp in degrees C at last zero interval

 

 

 

 

Box_Z_T

Float

4

Box temp in degrees C at last zeroing

 

 

 

 

PkPk_int

UI

2

Current peak to peak A/D counts from detector

 

 

 

 

PkPk

Float

4

Current peak to peak voltage from detector

 

 

 

 

Ave_PkPk

Float

4

Running average voltage from detector

 

 

 

 

Zero_PkPk

Float

4

Voltage measured at last zeroing

 

 

 

 

Noise

Float

4

Largest Change in running average

 

 

 

 

AU

Float

4

Current absorbance value

 

 

 

 

Ave_AU

Float

4

Running Average of absorbance level

 

 

 

 

Bench_PPM

Float

4

PPM in bench based on zone gas selection (uncorrected

 

 

 

for pressure and temperature)

 

 

 

 

STP_PPM

Float

4

PPM corrected to STP (1 atm, 25 deg C)

 

 

 

 

Bench_UML

Float

4

Micromoles/liter in bench (uncorrected0

 

 

 

 

Ambient_UML

Float

4

Micromoles/liter corrected to ambient pressure

 

 

 

 

 

 

76

Instruction 3015-4275

Image 84 Contents
AGM300 / ADM800 Ammonia Gas Monitoring System Rev MayInstruction Table of Contents How to Use This Manual IntroductionPlease Read this Manual Carefully Before Use Safety Precautions Hazard Symbols on Monitor Installation CategoryCleaning Altitude LimitFunctional Overview General DescriptionCommunication Options Suggested Location of Sampling PointsChiller Machine RoomStandard Accessories for a 4 Point System LEDAGM300 Installation Considerations Locating the MonitorInspection Installation AGM300 Mounting InstructionsScrew Locations AGM300 Mounting SpecsAGM300 Connecting Air Lines OverviewAGM300 Side View Inlet / Purge PortsConnecting Purge Line Connecting Exhaust LineConnecting Sample Intake Lines Tubing ConsiderationsAGM300 Interior Schematic Do not Apply Power Switches / FusesON/OFF / OUTAGM300 Electrical Wiring VAC InstallationAGM300 AC Input Power and Ground Connections AGM300 Connecting Communication Devices RS-485 ConnectorAmmonia Display Module ADM800 Connection Integrating with Building Management SystemsChanging Terminator Settings Personal Computer Refer to AGM300 PC Software SectionOUT Connecting to a Building Management System Multiple AGM’sRS-485 Connections Between AGM300’s Programming the AGM300 Using PC Software AGM300 PC SoftwareMultiple AGM300’s Connected to a Building Management System Unit Node Terminator Outout Building Management SystemSaving the AGM300 Program to a Disk or Drive To Open a Saved ProgramTo Send a Saved Program to an AGM300 Trend DataUSB Type Laptops Sealevel SYSTEMS, INCCurrent Loop Connector AGM300 Current Loop InterfacesOptional 4-20 mAdc Outputs ConnectionAGM300 Connecting External Alarms AGM300 Relay ConnectorLeak Spill Evacuate Fault Typical AGM300 Relay 1 and Relay 2 WiringADM800 Installation ADM800 Oblique PhotoADM800 Installation Considerations ADM800 Mounting Instructions ADM800 Mounting SpecsADM800 Interior Schematic Front Panel ChassisTo Monitors To HostADM800 Electrical Wiring ADM800 AC Input Power and Ground Connections ADM800 Communication Connections AGM300 NetworkIntegration with Building Management System Changing Terminator Switch Settings To MonitorADM800 Connecting External Alarms ADM800 Relay ConnectorADM800 Exterior Schematic ADM800 OperationADM800 Overview ADM800 Screen Displays System ScreenSetup Parameters Password ProtectionADM800 System Programming ADM800 Setup Programming Setting the ClockClock Setup Screen Navigating to the 1st ADM Setup Screen ContrastNumber of AGM Units AGM Baud RateAudible Alarm PasswordEnabling Building Management System Connection Building Management System Stop BitsNavigating to the 2nd ADM Setup Screen Setting Relay ParametersADM Setup Screen #2 AGM Node AddressAGM300 Setup Programming Navigating to the 1st and then 2nd AGM Setup ScreenScreen #1 AGM Setup Screen #1 AGM Setup Screen #2Number of Zones Installed Alarm Ack ModeLocation Zone Hold TimeRe-Zero Mode Loop 2 FactorAuto Zone ChangeNavigating to the 3rd AGM Setup Screen Service TimeoutAGM Setup #3 Screen ADM800 Zone Setup Programming Navigating to the 1st Zone Setup ScreenZone Setup Screen #1 DistanceAverage Temperature Navigating To the 2nd Zone Setup ScreenCurrent PPM Log IntervalRe-Setting the Peak PPM Value Spill LevelEvacuation Level General Operation Functional OverviewWorking with the Zone Screen Zone ScreenZone Screen Hold Mode Zone Screen Alarm ModeFault Conditions System Screen Alarm Mode Working with the System ScreenAGM Setup #1 Screen Alarm Log ScreenSystem Screen Fault Mode Working with Alarms Responding To AlarmsAlarm Summary Screen Alarm Detail Screen Alarm Detail ScreenAcknowledging Alarms Alarm Summary Screen Acknowledge ModeWorking with the Trend Screen Overview Log IntervalNavigating To the Trend Screen Trend ScreenWorking with System Faults Navigating to the Fault ScreenFault Screen Reset to Factory Default Settings Critical FaultsClearing System Faults Viewing Fault LogFault Log Screen Adjusting Calibration Factor Working with the Calibration ScreenNavigating to the Calibration Screen Calibration ScreenCalibration Procedure Working with the Diagnostic Screen Navigating to the Diagnostic ScreenDiagnostic Screen Overview Diagnostic ScreenInstruction Service Mode System Screen Service ModeInstruction Appendix AGM300 / ADM800 Maintenance Servicing Inline FilterServicing Air Lines Replacing FusesOptional Accessories Intake ManifoldsMA Interface Board ADM800 Logic Diagram RS-485 Communication Protocol Modbus RTU ProtocolAGM300 Modbus RTU Operation Protocol DetailsModbus Mode RTU only Key Comm Protocol ParametersSummary of Registers Data Type AbbreviationsRegister Name Number Type Description TIMREV Unused TIMUnused Minutes RS-485BAUD Service mode register to change this parameter Mode. do not Modify use zone hold register or MSB LSBZone Data Register 0x12xxh R/W 78 bytes Variable Type Length DescriptionGood. do not Modify Instruction Instruction Stpppm To Release the Zone Hold Use the Following Sequence of Steps Fault Log Register Register 0x1900, 0x1901h Bytes Instruction Point to current reading Time record for each of the 100 log points. The formatLast 100 log points 2 byes per point Warranty and Service WarrantyService Return Procedure Products Covered under Warranty and Service PolicyField Repair AGM300 Specifications ADM800 Specifications Instruction Headquarters Hunt Valley Circle, New Kensington, PA