Honeywell manual Specifications for 04973 Series, Specifications for 055919 Flow Chamber

Page 10

Specifications

2.Specifications

2.1Specifications for 04973 Series

Parameter

Description

 

 

Cell Constant

04973 Series: 0.01, 0.1, 1.0 and 10 as specified

 

 

Electrode Material

0.01 and 0.1 constant, titanium

 

1.0 and 10 constant, high density graphite with

 

Teflon sheath

Maximum Pressure Limit

1724 kPa (250 psig) at rated temperature

 

 

Maximum Continuous Temperature Limit

140°C (284°F) at rated pressure

 

For PVC wire:

 

105°C (221°F)

Materials of Construction

Cell Body: PES (polyethersulfone)

 

Electrodes: see ‘Electrode Material’

 

 

Cable Options

Leadwire: PVC insulated 18 gage cable

 

20 and 50 feet lengths available

 

Quick Disconnect Option

 

Universal Head (Aluminum)

 

 

Approvals

Manufactured to comply with ASME boiler and

 

pressure vessel code Section III, Div.1, UG-101

 

CRN #0F11607.5C

 

 

Insertion

3/4” NPT male, Schedule 40 and 80 pipes

 

 

Insertion Depth

3-1/2” (89 mm) for 1, 10 and 0.01 cell constants

 

from solution end of 3/4" MNPT

 

2-1/2” (64mm) for 0.1 cell constant

 

 

2.2 Specifications for 055919 Flow Chamber

Parameter

Description

 

 

Maximum Flow

2 gpm @ 40psig and atmospheric discharge

 

 

Maximum Pressure

200 psig @ 25°C

 

 

Maximum Temperature

140°C (284°F) at atmospheric pressure

 

 

Dimensions

See Figure 3-1

 

 

Materials of Construction

Polyethersulfone (PES)

 

 

4

Installation and Maintenance Manual

10/07

Image 10
Contents 70-82-25-17 10/07 Warranty/Remedy Copyright, Notices, and TrademarksRevision Notes AbstractReferences Contacts Contents Figures Overview IntroductionDescription Series Cells Only Preferred Specifications for 04973 Series SpecificationsSpecifications for 055919 Flow Chamber Installation General RequirementsIn-Line Type Mounting Flow-Type MountingDimension Drawings Torque RecommendationsDimension Drawing for 31079198 Flow Chamber Conductors Instrument Wiring for 4973 Cells with Integral Cable Electrical ConnectionsYellow Coax 20OR50FT.CABLELENGTH Model 4973 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4973 Cells with Quick Disconnect Cable Wiring Model 4973 with Quick Disconnect Cable to APT Troubleshooting MaintenanceTo Clean the Cell Check Conductivity SystemAir Entrapped in Cell Flow Channel Accessories and Spare Parts Extension Cables for Sensors with Quick Disconnect OptionPage Page Honeywell Field Solutions