Honeywell 4973 Series manual Air Entrapped in Cell Flow Channel

Page 22

Maintenance

5.5 Air Entrapped in Cell Flow Channel

If measurement errors appear for horizontal mountings of a 1 or 10 constant cell, it may be that air is entrapped in the cell flow channel. Take one of the following actions to eliminate this problem:

Increase flow to at least 1 gpm.

Rotate the cell mounting so that its flow channel faces the same direction as the pipeline flow exit.

Install the cell and/or flow chamber vertically.

16

Installation and Maintenance Manual

10/07

Image 22
Contents 70-82-25-17 10/07 Warranty/Remedy Copyright, Notices, and TrademarksRevision Notes AbstractReferences Contacts Contents Figures Overview IntroductionDescription Series Cells Only Preferred Specifications for 04973 Series SpecificationsSpecifications for 055919 Flow Chamber Installation General RequirementsIn-Line Type Mounting Flow-Type MountingDimension Drawings Torque RecommendationsDimension Drawing for 31079198 Flow Chamber Conductors Instrument Wiring for 4973 Cells with Integral Cable Electrical ConnectionsYellow Coax 20OR50FT.CABLELENGTH Model 4973 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4973 Cells with Quick Disconnect Cable Wiring Model 4973 with Quick Disconnect Cable to APT Troubleshooting MaintenanceTo Clean the Cell Check Conductivity SystemAir Entrapped in Cell Flow Channel Accessories and Spare Parts Extension Cables for Sensors with Quick Disconnect OptionPage Page Honeywell Field Solutions