Introduction

1. Introduction

1.1 Overview

The 4973 Type cells, Figure 1-1, have a rugged configuration for reliable, continuous measurements of electrolytic conductivity in industrial water processes at temperatures up to 140°C at 250 psig. They are applicable to such measurements as the effluent of distillation equipment; anion, cationic and mixed bed ion exchangers; monitoring and controlling of washing electronic components; plating rinse tank control; boiler water condensate; boiler hot well measurements and cooling tower blowdown, and many others.

Made of polyethersulfone (PES) construction for high-corrosion resistance, the cells are supplied with 0.01 and 0.1 cell constants having titanium electrodes, and 1.0 and 10.0 cell constants with high-density, graphite electrodes.

The 4973 Cells used with UDA2182 Series or APT2000/4000 Series are equipped with the choice of:

integral standard 20 or 50 foot lead

quick disconnect cable option

junction box type (universal) head with terminal connections for longer lead lengths.

For in-line applications, the 3/4” NPT male thread permits permanent installation in a pipe or tank; the cell may also be used as a laboratory dip-type cell for batch sampling.

For flow applications, the cell can be installed directly into a process stream as shown in Figure 1-2 by using a Flow Chamber or a 3/4” pipe tee as shown in Figure 1-3. These arrangements are designed to keep both the temperature compensator and cell in the main stream flow so that the cell will respond more quickly and accurately to process changes of both solution concentration and temperature.

A 316SS (P/N 31079198) and a PES (P/N 055919) Flow Chamber is available from Honeywell.

1.2 Description

All conductivity cells of the 4973 Type are suitable for use in both flow and insertion - type installations. They are one-piece molded units that cannot come apart and therefore have no replacement parts. Each has a 3/4” NPT thread.

The physical appearance of the cells is shown in Figure 1-1. The 0.01, 0.1, 1 and 10 cell constants are similar in construction with differences as noted below.

0.01 and 0.1 Cell Constants

The 0.01 cell differs from the others only by its outer electrode length of 2-3/4”. The temperature compensation sensor is located inside the inner electrode. The holes in the outer electrode provide passage for the solution being measured. The 0.1 cell is similar to the 0.01 type except that its outer electrode length is 2”. The temperature compensation sensor is located inside the inner electrode.

1 and 10 Cell Constants

The 1 and 10 cell constant types are similar but differ in the width of flow channel (that serves to conduct the solution being measured past the electrodes of the cell) as well as size and spacing of the electrodes. The electrodes are graphite, 1/4” D for the 1 cell constant and 1/8” D for the 10 cell constant. The temperature compensating sensor is integral with the cell body.

10/07

Installation and Maintenance Manual

1

Image 7
Contents 70-82-25-17 10/07 Copyright, Notices, and Trademarks Warranty/RemedyRevision Notes AbstractReferences Contacts Contents Figures Overview IntroductionDescription Series Cells Only Preferred Specifications for 04973 Series SpecificationsSpecifications for 055919 Flow Chamber General Requirements InstallationFlow-Type Mounting In-Line Type MountingTorque Recommendations Dimension DrawingsDimension Drawing for 31079198 Flow Chamber Conductors Electrical Connections Instrument Wiring for 4973 Cells with Integral CableYellow Coax 20OR50FT.CABLELENGTH Model 4973 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4973 Cells with Quick Disconnect Cable Wiring Model 4973 with Quick Disconnect Cable to APT Check Conductivity System MaintenanceTo Clean the Cell TroubleshootingAir Entrapped in Cell Flow Channel Extension Cables for Sensors with Quick Disconnect Option Accessories and Spare PartsPage Page Honeywell Field Solutions