Honeywell 4973 Series manual Dimension Drawing for 31079198 Flow Chamber

Page 14

Installation

4.25

8-32 UNC – 2B 250 DEEP

107.9

 

 

1.0

.875

25.4

.50

22.22

12.7

 

 

1.5

 

38.1

3.75

 

95.25

 

4.0

 

101.6

1/8 NPT

 

4.5

 

114.3

3/4 NPT

Figure 3-2 Dimension Drawing for 31079198 Flow Chamber

0.593" Dia. (15mm)

 

"X"

5.875"

See Table

(149.2mm)

Table I

Dim "X"

Inch

mm

0.01

3.3

84

0.1

2.6

66

1.0

3.5

89

10

3.5

89

3/4" NPT

1.13" Hex (29mm)

¾” female NPT for user's flexible electrical conduit

connection. For insertion or removal of cell, disconnect conduit connections.

3/4” (76mm)

Four Point Terminal Board for lead wire connections. Each #6-32 screw terminal will accomodate one

#12 or smaller AWG wire

NOTE: For existing users with conduit, a ¾” x ½” adapter bushing will be required to use existing conduit.

Figure 3-3 Outline and Dimensions for Conductivity Cell with Universal Head

8

Installation and Maintenance Manual

10/07

Image 14
Contents 70-82-25-17 10/07 Warranty/Remedy Copyright, Notices, and TrademarksReferences AbstractRevision Notes Contacts Contents Figures Description IntroductionOverview Series Cells Only Preferred Specifications for 055919 Flow Chamber SpecificationsSpecifications for 04973 Series Installation General RequirementsIn-Line Type Mounting Flow-Type MountingDimension Drawings Torque RecommendationsDimension Drawing for 31079198 Flow Chamber Conductors Instrument Wiring for 4973 Cells with Integral Cable Electrical ConnectionsYellow Coax 20OR50FT.CABLELENGTH Model 4973 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4973 Cells with Quick Disconnect Cable Wiring Model 4973 with Quick Disconnect Cable to APT Troubleshooting MaintenanceTo Clean the Cell Check Conductivity SystemAir Entrapped in Cell Flow Channel Accessories and Spare Parts Extension Cables for Sensors with Quick Disconnect OptionPage Page Honeywell Field Solutions