Installation
3.5 Torque Recommendations
For inserting a cell in metal fittings or bushings - 40
3.6 Dimension Drawings
1 1/ 8" | 1 1/2" | Flow Out |
|
3/4" NPT (Female) |
| ||
(38mm) | Flow Chamber | ||
Hexagon | (38mm) | 1 1/2" | |
|
| Flow In | |
|
| (38mm) | 3/4" NPT |
1 1/2" |
| Octagon | |
|
| (Male) | |
(38mm) |
| CELL | IN |
|
|
83/4"
(222mm)
14 1/2" max (368mm)
3/4" Fitting
Allow 4 1/8" (105mm) for re moval of cell
Notes:
1.Mount cell and flow chamber horizontally as shown above with flow exit up to eliminate possible air gap around cell body.
2.If cell and flow chamber must be mounted vertically, attach a short length of tubing to flow
exit as shown below and form a trap to ensure filling of flow chamber, especially at low flow.
CELL
IN
2" min. (51mm)
a/n23342
Figure 3-1 Dimension Drawing for 055919 Flow Chamber
10/07 | Installation and Maintenance Manual | 7 |