Fluke 7000DP, 6000D manual Load resistance, Pulse width

Page 2

Impulse 6000D/7000DP Users

Manual Supplement

Change #1, 46470, 46697, 47714, 48007, 49680

Replace pages 31 through 35, with the following:

Defibrillator Analyzer Specifications

Energy Output Measurement

Compatible Defibrillator Waveshapes

Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic

Note

AC Pulsed Biphasic waveform has not been approved in the United States.

Autoranged Measurement

0.1 to 600 J

Accuracy

 

0.1 to 360 J

±(1 % of reading + 0.1 J)

360 to 600 J

±(1 % of reading + 0.1 J), typical

Note

For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges.

Load resistance

 

Resistance

50 Ω

Accuracy

±1 %, non-inductive (<2 μH)

Pulse trigger level

20 V

Pulse width

 

Range

1.0 to 50.0 ms

4/08

1

Image 2
Contents Manual Supplement Load resistance Pulse widthCurrent Interphase delay biphasic and pulsed biphasicVoltage Tilt biphasic and pulsed biphasicAutomated Defibrillator Test ECG Waves Scope OutputCharge Time Measurement ECG General ECG AmplitudesECG on Defibrillator Input Load ECG Normal SinusECG High Level Output BNC Jack ECG Performance WavesTransvenous Pacer Pulse Simulation Wave DetectionNoise Immunity ±10 % of setting + 0.2 mV Change #2 Setting the ECG Reference Lead Change #3