Fluke 7000DP, 6000D manual ±10 % of setting + 0.2 mV

Page 8

Impulse 6000D/7000DP Users

Manual Supplement

Accuracy

±(10 % of setting + 0.2 mV)

Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)

Lead II reference

 

Lead #

 

I

 

II

 

III

 

V1

 

V2

 

V3

 

V4

 

V5

 

V6

 

Ref. Amp.

 

67 %

100 %

33 %

67 %

67 %

67 %

67 %

67 %

67 %

Lead I reference

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lead #

 

I

 

II

 

III

 

V1

 

V2

 

V3

 

V4

 

V5

 

V6

 

Ref. Amp.

 

100 %

 

150 %

 

50 %

 

100 %

 

100 %

 

100 %

 

100 %

 

100 %

 

100 %

Arrhythmia Selections

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)

 

 

 

 

 

 

 

 

Demand

 

 

 

30 to 360 (by 1) BPM

 

 

 

 

 

 

 

 

Asynchronous

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Non-Capture

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Non-Function

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Threshold (Interactive pacing simulation only)10 to 250 (by 10) mA

Supraventricular

Atrial Fibrillation Coarse

Atrial Fibrillation fine

Atrial Flutter

Sinus Arrhythmia

Missed Beat

Atrial Tachycardia

4/08

7

Image 8
Contents Manual Supplement Load resistance Pulse widthInterphase delay biphasic and pulsed biphasic VoltageCurrent Tilt biphasic and pulsed biphasicAutomated Defibrillator Test ECG Waves Scope OutputCharge Time Measurement ECG General ECG AmplitudesECG Normal Sinus ECG High Level Output BNC JackECG on Defibrillator Input Load ECG Performance WavesTransvenous Pacer Pulse Simulation Wave DetectionNoise Immunity ±10 % of setting + 0.2 mV Change #2 Setting the ECG Reference Lead Change #3