Fluke 6000D, 7000DP manual Wave Detection, Noise Immunity, Transvenous Pacer Pulse Simulation

Page 7

Manual Supplement

Impulse 6000D/7000DP Users

Sine waves

0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz

Pulse

30 and 60 BPM, 60 ms pulse width

R-Wave Detection

 

Waveform

Haver-triangle

Amplitude

0.05 to 0.45 (by 0.05) mV

 

0.5 to 5.0 (by 0.5) mV

Rate

30, 60, 80, 120, 200, and 250 BPM

Widths

8, 10, 12 ms, and 20 to 200 (by 10) ms

Accuracy

±(1 % setting + 1 ms)

Noise Immunity

 

Wave

Sine

Line Frequency

50 or 60 Hz (± 0.5 Hz)

Amplitude

 

Range

0.0 to 10.0 (by 0.5) mV

Accuracy

±5 %

Transvenous Pacer Pulse Simulation

 

Widths

 

Range

0.1, 0.2, 0.5, 1.0, and 2.0 ms

Accuracy

±5 % of setting

Amplitude

 

Range

0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and

 

±700 mV

6

4/08

Image 7
Contents Manual Supplement Pulse width Load resistanceTilt biphasic and pulsed biphasic Interphase delay biphasic and pulsed biphasicVoltage CurrentCharge Time Measurement Scope OutputAutomated Defibrillator Test ECG Waves ECG Amplitudes ECG GeneralECG Performance Waves ECG Normal SinusECG High Level Output BNC Jack ECG on Defibrillator Input LoadNoise Immunity Wave DetectionTransvenous Pacer Pulse Simulation ±10 % of setting + 0.2 mV Change #2 Setting the ECG Reference Lead Change #3