Fluke 6000D Voltage, Current, Tilt biphasic and pulsed biphasic, Frequency pulsed biphasic only

Page 3

Manual Supplement

Impulse 6000D/7000DP Users

Accuracy

±0.1 ms

Voltage

 

Range

20 to 5000 V

Accuracy

±(1 % of reading + 2 V)

Current

 

Range

0.4 to 100.0 A

Accuracy

±(1 % of reading + 0.1 A)

Tilt (biphasic and pulsed biphasic)

 

Range

1 % to 99 %

Accuracy

±1 digit

Interphase delay (biphasic and pulsed biphasic)

Range

0.1 ms to 9.9 ms

Accuracy

±0.1 ms

Frequency (pulsed biphasic only)

 

Range

2000 Hz to 8000 Hz

Accuracy

±1 % of reading

Duty cycle (pulse biphasic only)

 

Range

1 % to 99 %

Accuracy

±1 digit

Sample rate

250 kHz (4 μs sample)

Maximum Average Power

12 W, equivalent to 10 defib pulses of 360 J every 5 minutes

2

8/08

Image 3
Contents Manual Supplement Pulse width Load resistanceTilt biphasic and pulsed biphasic Interphase delay biphasic and pulsed biphasicVoltage CurrentScope Output Charge Time MeasurementAutomated Defibrillator Test ECG Waves ECG Amplitudes ECG GeneralECG Performance Waves ECG Normal SinusECG High Level Output BNC Jack ECG on Defibrillator Input LoadWave Detection Noise ImmunityTransvenous Pacer Pulse Simulation ±10 % of setting + 0.2 mV Change #2 Setting the ECG Reference Lead Change #3