Fluke 7000DP ECG Normal Sinus, ECG High Level Output BNC Jack, ECG on Defibrillator Input Load

Page 6

Impulse 6000D/7000DP Users

 

 

 

 

 

 

 

 

Manual Supplement

Normal Sinus waves

 

 

 

 

 

 

 

 

 

 

 

 

Lead #

 

I

II

III

V1

 

V2

V3

V4

V5

V6

 

 

Ref. Amp.

 

70 %

100 %

30 %

24 %

 

48 %

100 %

120 %

112 %

80 %

 

Lead I reference

 

 

 

 

 

 

 

 

 

 

 

Performance waves and R wave detection

 

 

 

 

 

 

 

 

 

Lead #

 

I

II

III

V1

 

V2

V3

V4

V5

V6

 

 

Ref. Amp.

 

100 %

150 %

50 %

100 %

 

100 %

100 %

100 %

100 %

100 %

 

Normal Sinus waves

 

 

 

 

 

 

 

 

 

 

 

 

Lead #

 

I

II

III

V1

 

V2

V3

V4

V5

V6

 

 

Ref. Amp.

 

100 %

150 %

50 %

24 %

 

48 %

100 %

120 %

112 %

80 %

 

ECG Normal

Sinus

 

 

 

 

 

 

 

 

 

 

 

 

Rates

 

10 to 360 (by 1) BPM

 

 

 

 

 

 

ECG High Level Output (BNC Jack)

 

 

 

 

 

 

 

 

 

 

Amplitude

 

 

 

 

 

 

 

 

 

 

 

Range

 

0.5 V per mV of reference lead setting

 

 

 

 

Accuracy

 

±5 %

 

 

 

 

 

 

 

 

Output Impedance

 

50 Ω

 

 

 

 

 

 

 

 

ECG on Defibrillator Input Load

 

 

 

 

 

 

 

 

 

 

Same as the LEAD II amplitude but limited to ±4 mV

 

 

 

 

 

 

 

 

ECG Performance Waves

 

 

 

 

 

 

 

 

 

 

 

Square wave

 

2.0 and 0.125 Hz

 

 

 

 

 

 

Triangular wave

 

2.0 and 2.5 Hz

 

 

 

 

 

 

8/08

5

Image 6
Contents Manual Supplement Load resistance Pulse widthCurrent Interphase delay biphasic and pulsed biphasicVoltage Tilt biphasic and pulsed biphasicScope Output Charge Time MeasurementAutomated Defibrillator Test ECG Waves ECG General ECG AmplitudesECG on Defibrillator Input Load ECG Normal SinusECG High Level Output BNC Jack ECG Performance WavesWave Detection Noise ImmunityTransvenous Pacer Pulse Simulation ±10 % of setting + 0.2 mV Change #2 Setting the ECG Reference Lead Change #3