Fluke 6000D, 7000DP manual ECG General, ECG Amplitudes

Page 5

Manual Supplement

Impulse 6000D/7000DP Users

Ventricular Fibrillation

Coarse and fine

Monomorphic Ventricular Tachycardia

120 to 300 (by 5) BPM

Polymorphic Ventricular Tachycardia

5 types

Asystole

Flat line

ECG Waves

 

ECG General

 

Lead configuration

12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs

Lead to lead impedance

1000 Ω (nominal)

Rate accuracy

±1 % of nominal

ECG Amplitudes

 

Reference Lead

Selectable, Lead II (default) or Lead I

Settings

0.05 to 0.45 (by 0.05) mV

0.5to 5.0 (by 0.5) mV

Accuracy (All Performance waves and Normal Sinus R waves)

Lead II

 

 

 

±2 %

 

 

 

 

 

 

All other leads

 

 

 

±5 %

 

 

 

 

 

 

Defib paddles

 

 

 

±5 %

 

 

 

 

 

 

Amplitude of ECG signals relative to amplitude setting (in percent)

 

 

 

 

 

Lead II reference

 

 

 

 

 

 

 

 

 

 

Performance waves and R wave detection

 

 

 

 

 

 

 

 

Lead #

I

II

 

III

V1

V2

V3

V4

V5

V6

 

Ref. Amp.

70 %

100 %

 

30 %

100 %

100 %

100 %

100 %

100 %

100 %

4

3/09

Image 5
Contents Manual Supplement Pulse width Load resistanceVoltage Interphase delay biphasic and pulsed biphasicCurrent Tilt biphasic and pulsed biphasicAutomated Defibrillator Test ECG Waves Scope OutputCharge Time Measurement ECG Amplitudes ECG GeneralECG High Level Output BNC Jack ECG Normal SinusECG on Defibrillator Input Load ECG Performance WavesTransvenous Pacer Pulse Simulation Wave DetectionNoise Immunity ±10 % of setting + 0.2 mV Change #2 Setting the ECG Reference Lead Change #3