Liebherr 7083 247-00 External alarm, Floating alarm output, RS485 interface, Rx- / Tx, Rx+ / Tx+

Page 12

External alarm

We recommend connecting the appliance to an external alarm device.

There are various connection options at the back of the appliance.

The appliance may only be connected to an external alarm device by trained personnel.

Floating alarm output

These three contacts can be used to connect the appliance to an optical or acoustic alarm device.

The connection is designed for a maximum of 42 V/8 A DC from a safety extra-low voltage (SELV) source (minimum current: 150 mA).

Important

When supplying mains voltage to the floating alarm contact, the technical safety requirements of standard EN 60335 will not be satisfied.

N.O

Optional temperature sensor connection

Floating alarm output

Connection to RS485 interface

Terminating resistor

When connecting several appli- ances via the RS485 interface, the terminating resistor must remain on the last appliance.

Remove the terminating resis- tors from the appliances in between.

Alarm output

Connection for a visual warning light or an acoustic alarm signal.

N.C

Operating light

Connection for a control lamp to indicate that the appliance is in normal mode.

COM

External power supply unit 42 V/8 A DC maximum. Minimum current: 150 mA.

RS485 interface

Rx- / Tx-

Note

The connectors are secured with screws. To remove the connectors, undo the left and right screws.

Send/Receive data cable (negative pole)

Rx+ / Tx+

Send/Receive data cable (positive pole)

GND

Earth cable

24

Image 12
Contents 7083 LKUvAppliance dimensions Disposal notesRange of appliance use Description of the applianceSafety instructions and warnings Setting upSwitching the appliance on and off Setting the temperatureAudible warning signal Door open alarmSetting the delay time for the door open alarm Deactivating the audible warning signal functionAudible warning signal settings Display =Alarm messages Alarm testAdjusting the alarm parameters Resetting the recorded temperature progres- sion Example of an alarm querySituation Haccp flashes in the display Display = Calibrating the control sensor standard sensor for Calibrating the product sensorActivating the sensor Resetting the parameters to factory settings Setting the real time clockChanging the network address Safety lock Setting the display indication for the defrost phaseCooling DefrostingCleaning Do not use steam cleaners because Risk of injury and damageOpening for external temperature sensor MalfunctionsExternal alarm Floating alarm outputRS485 interface Rx- / TxInstallation dimensions mm Changing over door hingesVersion Unscrew the hinge bracket