2.2.8. Electrical Requirements
The CMA-180 comes standard factory, wired for 3-phase operation. Check the electrical data plate to confirm this. If the account requires 1 phase, refer to “Electrical Requirements”, Section 1-A, for proper wiring instruction for conversion. Also check the wiring diagram to properly wire the terminal power block, tank heater, and booster heater for 1 phase (or 1B diagram below).
SINGLE PHASE POWER
NOTE: 80 amp service is required when CMA-180 is wired for single-phase with the booster heater. See “Electrical Requirements” section for DUAL 1-phase power supply. Circuit breaker requirement (1) 50 amp & 30 amp.
Figure 1-B
| | Single Phase | | Single Phase |
| | | | | |
| | Rinse Booster Heater | | Wash Tank Heater | | |
| | | | |
| | | | |
| | | | | | | |
1. | Install booster heater | 1. | Install wash heater bus |
| | | wires and bus bars as | | bars as shown above. |
2. | Connect wire #59 to | 2. | Connect wire #15 to |
| | | heater L1. | | heater L1. |
3. | Connect wire #57 to | 3. | Connect wire #17 to |
| | | heater L2. | | heater L2. |
| | | | | | |
MODEL CMA-180 INSTALLATION & OPERATION Rev. 2.03 | | Page 12 |