General Information
Vertical System: Analog Channels
*Denotes Warranted Specifications, all others are typical. Specifications are valid after a
Vertical System: Analog Channels
Analog channels | 54621A/21D, 54622A/22D: Ch1 and 2 simultaneous acquisition |
| 54624A: Ch 1, 2, 3, and 4 simultaneous acquisition |
Bandwidth | 54621A/22D: dc to 60 MHz |
| 54622A/22D/24A: dc to 100 MHz |
ac coupled | 54621A/21D: 3.5 Hz to 60 MHz |
| 54622A/22D/24A: 3.5 Hz to 100 MHz |
Calculated rise time | 54621A/22D: ~5.8 ns |
(= 0.35/bandwidth) | 54622A/22D/24A: ~3.5 ns |
Single Shot Bandwidth | 50 MHz |
Range1 | 1 mV/div to 5 V/div |
Maximum Input | CAT I 300 Vrms, 400 Vpk |
| CAT II 100 Vrms, 400 Vpk |
| with 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk |
Offset Range | ±5 V on ranges <10 mV/div |
| ±25 V on ranges 10 mV/div to 199 mV/div |
| ±100 V on ranges ≥200 mV/div |
Dynamic Range | Lesser of ±8 div or ±32 V |
Input Resistance | 1 MΩ ±1% |
Input Capacitance | ~ 14 pF |
Coupling | ac, dc, ground |
BW Limit | ~ 20 MHz selectable |
dc to 20 MHz > 40 dB | |
(with channels at same V/div) | 20 MHz to max bandwidth > 30 dB |
Probes | 10:1 10074C shipped standard for each analog channel |
Probe ID (Agilent/HP & | Auto probe sense |
Tek Compatible) |
|
11 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.