Calibrating and Adjusting
To adjust the power supply
2 Locate the
The power supply voltages are not labeled on the system board. See
Figure 4-1 for the locations.
9
9
9
9
Low Voltage Power Supply Voltages (on the bottom of the oscilloscope)
3Make sure that the voltage measurements are within the tolerances listed in Table
Power Supply Voltage Tolerances
Supply Voltage | Tolerance |
+5.1 V | ±250 mV (4.85 V to 5.35 V) |
±156 mV | |
+15.75 V | +1.260 V, |
+3.3 V | ±100 mV (+3.20 V to +3.40 V) |
+8.2 V | ±82 mV (+8.12 V to +8.28 V) |
| measure at J4 on the power supply board |