Agilent Technologies 54622D Agilent 54621A/21D Performance Test Record, Agilent 54621A/54621D

Page 62

Agilent 54621A/21D Performance Test Record

Agilent 54621A/54621D

Serial No. ______________________________________

Test by _____________________________

Test Interval ____________________________________

Work Order No. ______________________

Recommended Next Testing ________________________

Temperature ____________

 

 

 

 

 

 

 

 

Threshold

Specification

Limits

Ch D7-D0

Ch D15-D8

 

 

Accuracy Test

5 V - 250 mV

4.750 V

________

________

 

 

(100 mV + 3% of

5 V + 250 mV

5.250 V

________

________

 

 

threshold setting)

-5 V - 250 mV

-5.250 V

________

________

 

 

 

-5 V + 250 mV

-4.750 V

________

________

 

 

 

0 V - 100 mV

-100 mV

________

________

 

 

 

0 V + 100 mV

100 mV

________

________

 

 

 

 

 

 

 

 

Voltage Measurement Accuracy

 

 

 

 

 

Range

Power Supply Setting

Test Limits

Channel 1

Channel 2

Channel 3

Channel 4

5 V/Div

35 V

34.04 V to 35.96 V

________

________

________

________

2 V/Div

14 V

13.616 V to 14.384 V

________

________

________

________

1 V/Div

7 V

6.808 V to 7.192 V

________

________

________

________

500 mV/Div

3.5 V

3.404 V to 3.596 V

________

________

________

________

200 mV/Div

1.4 V

1.3616 V to 1.4384 V

________

________

________

________

100 mV/Div

700 mV

680.8 mV to 719.2 mV

________

________

________

________

50 mV/Div

350 mV

340.4 mV to 359.6 mV

________

________

________

________

20 mV/Div

140 mV

136.16 mV to 143.84 mV

________

________

________

________

10 mV/Div

70 mV

68.08 mV to 71.92 mV

________

________

________

________

5 mV/Div

35 mV

34.04 mV to 35.96 mV

________

________

________

________

2 mV/Div

14 mV

13.616 mV to 14.384 mV

________

________

________

________

1 mV/Div

7 mV

6.616 mV to 7.384 mV

________

________

________

________

 

 

 

 

 

 

 

Bandwidth

 

Test Limits

Channel 1

Channel 2

 

 

 

 

3 dB at 60 MHz

________

________

 

 

 

 

 

 

 

 

Horizontal t and 1/t Accuracy

 

 

 

 

 

 

Generator Setting

Test Limits

Results

 

 

 

Frequency

10 kHz

9.98 kHz to 10.02 kHz

________

 

 

 

Period

100 s

99.79 s to 100.2 s

________

 

 

 

Frequency

1 MHz

997.9 kHz to 1.002 MHz

________

 

 

 

Period

1 s

997.9 ns to 1.002 s

________

 

 

 

Frequency

50 MHz

49.77 MHz to 50.23 MHz

________

 

 

 

Period

20 ns

19.91 ns to 20.09 ns

________

 

 

 

 

 

 

 

 

 

 

Trigger Sensitivity

 

Test Limits

Channel 1

Channel 2

 

 

 

Internal trigger

25 MHz at 0.35 division

________

________

 

 

 

 

60 MHz at 0.35 division

________

________

 

 

 

 

 

Ext

 

 

 

 

External trigger

25 MHz at <75 mVp-p

________

 

 

 

 

 

60 MHZ at <150 mVp-p

________

 

 

 

3-21

Image 62
Contents Agilent 54621D/22D Mixed-Signal Oscilloscope Softkeys extend the functionality of command keys Display shows current input signalsHorizontal Controls select sweep speed and delay parameters UtilitiesThis Book Contents Contents Solving General Problems with the OscilloscopesTroubleshooting the Oscilloscope General Information General Information To inspect package contents General Information Package contents for 54620-Series Oscilloscopes To inspect options and accessories Option DescriptionAccessories available Model DescriptionPower Cords Plug Type Cable Part NumberAcquisition Digital Channels on 54621D and 54622D only Performance CharacteristicsAcquisition Analog Channels Vertical System Analog Channels Probe ID Agilent/HP Auto probe sense Tek CompatibleVertical System Digital Channels 54621D and 54622D only Accuracy = ±2.0%80 mV + 0.4%80 mV = ±1.92 mVHorizontal Sample rate of 200/400 MSa/sTrigger System CanExternal EXT Triggering Analog Channel TriggeringDigital D15 D0 Channel Triggering 54621D and 5462 2D Measurement Features Display SystemFFT StorageGeneral Characteristics Power RequirementsPreparing the Oscilloscope for Use Preparing the Oscilloscope for Use Setting up the Oscilloscope To adjust the handle To adjust the handleTo power-on the oscilloscope To power-on the oscilloscopePress the power switch Intensity control To adjust the display intensityTo adjust the display intensity To connect the oscilloscope probes To connect the oscilloscope probesTo compensate your probe Press AutoscaleTo use the digital probes mixed-signal oscilloscope only To use the digital probes mixed-signal oscilloscope onlyGrabber Signal Lead Ground Lead Grabber Signals Ground To connect a printer To connect a printerTo connect an RS-232 cable To verify basic oscilloscope operation To verify basic oscilloscope operationPress the Autoscale key on the front panel Verifying Basic Oscilloscope OperationGetting started using the oscilloscope interface To verify basic oscilloscope operation Using Quick Help Preparing the Oscilloscope for Use Loading a language from floppy disk Loading a language from floppy diskCleaning the oscilloscope Disconnect power from the instrumentTesting Performance Testing Performance If a performance test failsList of Test Equipment Equipment Critical SpecificationsList of Test Equipment Part NumberMaterials Required to Construct the Test Connectors To construct the test connectorTo construct the test connector Constructing the 8-by-2 ConnectorTo test the 54621D/22D Oscilloscope digital channels To test the 54621D/22D Oscilloscope digital channelsPress the Autoscale key To verify digital channel threshold accuracy To verify digital channel threshold accuracyMethod 1 Using the Test Connector Method 2 Using the Test Fixture Threshold Accuracy Voltage Test Settings Threshold voltage setting DC offset voltage setting LimitsTo verify voltage measurement accuracy Set up the oscilloscopeEquipment Critical Specifications Recommended Model/Part To verify voltage measurement accuracySettings Used to Verify Voltage Measurement Accuracy MV/Div 616 mV 384 mVUsing a Blocking Capacitor to Reduce Noise Use a Blocking Capacitor to Reduce NoiseEquipment Required to Verify Bandwidth To verify bandwidthTo verify bandwidth Change the time base to 5 ns/div Calculate the response using this formulaSignal Generator Frequency Setting Selected Channel 54621A/21D 54622A/22D 54624ATo verify horizontal ∆t and 1/∆t accuracy To verify horizontal Dt and 1/Dt accuracy54621A and 54621D only Measure the following54622A/22D/24A only Equipment Required to Verify Trigger Sensitivity To verify trigger sensitivityTo verify trigger sensitivity Test Internal Trigger Sensitivity Test External Trigger Sensitivity Agilent 54622A/22D/24A Performance Test Record Agilent 54621A/21D Performance Test Record Agilent 54621A/54621DCalibrating and Adjusting Calibrating and Adjusting Hazardous Voltages To adjust the power supply To adjust the power supplyPrepare the oscilloscope for the voltage adjustment Digital multimeterPower Supply Voltage Tolerances Supply Voltage Tolerance+3.3V Adjustment To perform User Cal To perform User CalWhen to Adjust the Oscilloscope Display To adjust the oscilloscope displayTo adjust the oscilloscope display Do not adjust Sub Bri or Cont controls96,= Troubleshooting Troubleshooting Read All Cautions and WarningsSolving General Problems with the Oscilloscopes If there is no trace displayIf the trace display is unusual or unexpected If the trace display is unusual or unexpectedIf you cannot see a channel If you cannot see a channelTroubleshooting the Oscilloscope Equipment Required to Troubleshoot the OscilloscopeTo construct your own dummy load Connector Load Amps Resistor WattsLocation of the Low Voltage Power Supply Connectors To construct your own dummy loadTo check out the oscilloscope Connect the power cord, then turn on the oscilloscopeTo check out the oscilloscope Signal Name Frequency Pulse Width Voltage Display Signals on the System Board All OscilloscopesLow Voltage Power Supply Tolerances Location of the Fan Connector Is the fan running?Low Voltage Power Supply Test Points To check the Low Voltage Power SupplyTo check the Low Voltage Power Supply If the Fuse is Blown Beware of Hazardous VoltagesTo run the internal self-tests To run the internal self-testsReplacing Assemblies Replacing Assemblies Avoid Electrical Shock Removing the cabinet To remove the cabinetTo remove the cabinet Removing the fan To remove the fanTo remove the fan Removing the floppy drive To remove the floppy driveTo remove the floppy drive Removing the intensity shaft and front panel To remove the front panelTo remove the front panel Removing the display To remove the displayTo remove the display Unhooking the power switch shaft Preparing to remove the system board To remove the system boardTo remove the system board Removing the system board Disconnecting ground wire and ribbon cable To remove the power supplyTo remove the power supply Unhooking the power switch shaft Removing the power supply To remove the keyboard assembly To remove the keyboard assemblyLift the keyboard assembly straight out Removing the keyboard assemblyTo remove the handle To remove the handleReplaceable Parts Replaceable Parts To order a replacement part To order a replacement partExploded View of the 54621A/22A/24A/21D/22D Oscilloscope To order a replacement part Replaceable Parts Reference Agilent Part Qty Description Designator NumberMP1 Input Cable 54621D/22D only Option Accessory pouch and front-panel coverProbes Product Name Manufacturer’s NameManufacturer’s Address Model NumbersSound Pressure Level SafetyProduct Regulations Regulatory Information for CanadaSafety Document Warranty Agilent Technologies, Inc