Agilent 54622A/22D/24A Performance Test Record
Agilent 54622A/54622D/54624A
Serial No. ______________________________________ | Test by _____________________________ | |||||
Test Interval ____________________________________ | Work Order No. ______________________ | |||||
Recommended Next Testing ________________________ | Temperature ____________ |
| ||||
|
|
|
|
|
|
|
Threshold | Specification | Limits | Ch | Ch |
|
|
Accuracy Test | 5 V - 250 mV | 4.750 V | ________ | ________ |
|
|
(100 mV + 3% of | 5 V + 250 mV | 5.250 V | ________ | ________ |
|
|
threshold setting) | ________ | ________ |
|
| ||
| ________ | ________ |
|
| ||
| 0 V - 100 mV | ________ | ________ |
|
| |
| 0 V + 100 mV | 100 mV | ________ | ________ |
|
|
|
|
|
|
|
| |
Voltage Measurement Accuracy |
|
|
|
|
| |
Range | Power Supply Setting | Test Limits | Channel 1 | Channel 2 | Channel 3 | Channel 4 |
5 V/Div | 35 V | 34.04 V to 35.96 V | ________ | ________ | ________ | ________ |
2 V/Div | 14 V | 13.616 V to 14.384 V | ________ | ________ | ________ | ________ |
1 V/Div | 7 V | 6.808 V to 7.192 V | ________ | ________ | ________ | ________ |
500 mV/Div | 3.5 V | 3.404 V to 3.596 V | ________ | ________ | ________ | ________ |
200 mV/Div | 1.4 V | 1.3616 V to 1.4384 V | ________ | ________ | ________ | ________ |
100 mV/Div | 700 mV | 680.8 mV to 719.2 mV | ________ | ________ | ________ | ________ |
50 mV/Div | 350 mV | 340.4 mV to 359.6 mV | ________ | ________ | ________ | ________ |
20 mV/Div | 140 mV | 136.16 mV to 143.84 mV | ________ | ________ | ________ | ________ |
10 mV/Div | 70 mV | 68.08 mV to 71.92 mV | ________ | ________ | ________ | ________ |
5 mV/Div | 35 mV | 34.04 mV to 35.96 mV | ________ | ________ | ________ | ________ |
2 mV/Div | 14 mV | 13.616 mV to 14.384 mV | ________ | ________ | ________ | ________ |
1 mV/Div | 7 mV | 6.616 mV to 7.384 mV | ________ | ________ | ________ | ________ |
|
|
|
|
|
|
|
Bandwidth |
| Test Limits | Channel 1 | Channel 2 | Channel 3 | Channel 4 |
|
| 3 dB at 100 MHz | ________ | ________ | ________ | ________ |
|
|
|
|
|
| |
Horizontal ∆t and 1/∆t Accuracy |
|
|
|
|
| |
| Generator Setting | Test Limits | Results |
|
|
|
Frequency | 10 kHz | 9.98 kHz to 10.02 kHz | ________ |
|
|
|
Period | 100 ∝s | 99.79 ∝s to 100.2 ∝s | ________ |
|
|
|
Frequency | 1 MHz | 997.9 kHz to 1.002 MHz | ________ |
|
|
|
Period | 1 ∝s | 997.9 ns to 1.002 ∝s | ________ |
|
|
|
Frequency | 100 MHz | 99.10 MHz to 100.9 MHz | ________ |
|
|
|
Period | 10 ns | 9.91 ns to 10.09 ns | ________ |
|
|
|
|
|
|
|
|
|
|
Trigger Sensitivity |
| Test Limits | Channel 1 | Channel 2 | Channel 3 | Channel 4 |
| Internal trigger | 25 MHz at 0.35 division | ________ | ________ | ________ | ________ |
|
| 100 MHz at 0.35 division | ________ | ________ | ________ | _______ |
|
|
| Ext |
|
|
|
| External trigger | 25 MHz at <75 | ________ |
|
|
|
|
| 100 MHz at <150 | ________ |
|
|
|