TROUBLESHOOTING GUIDE

TROUBLESHOOTING GUIDE

Trouble Shooting Guide

In the event of an error code, unplug the Attest 290 Auto-reader and plug it back in to initiate the self- diagnostic test. If error code re-occurs, call for repair or replacement.

The display panel will indicate malfunctions detected by a specific error code. Refer to the following:

Code

Problem

Solution

C1 Unit Warm-up Caution Incubation block not at temperature (RRBI readings are disabled until incubator block reaches 60ºC +/-2ºC).

Allow a 30 minute warm-up after Attest 290 Auto-reader is plugged in.

C2 RRBI has been removed before incubation Replace RRBI in proper well within 10 seconds. complete. All three lights for affected RRBI

will light and alarm will sound if active.

E1

Temperature Control Error

 

This error will occur if system is no

longer able to control the temperature of the incubator block (60ºC +/-2ºC).

Call for repair or replacement.

E3

Memory Error.

Call for repair or replacement.

E4

Analog-Digital Conversion Error.

Call for repair or replacement.

E7

Communications Error

Call for repair or replacement.

 

This error will occur if internal

 

 

communications have failed.

 

E8

Bulb Error

 

This error occurs if the system detects

 

that the bulb is too dim to properly read

 

RRBI results.

Call for repair or replacement.

E9

Transport Error

 

This error will occur if the system detects

 

a motor failure or positioning problem.

Call for repair or replacement.

3M © 2003 78-8078-8977-5