INSTRUCTIONS FOR USE
Positive Control
Crush, tap and incubate at least one
It is good practice to use a positive control each day a sterilized RRBI is processed. This helps ensure:
-correct incubation temperatures are met.
-viability of spores has not been altered due to improper storage temperature, humidity, or proximity to chemicals
-capability of media to promote rapid growth, and
-proper functioning of Attest 290
Interpretation of Results
Each incubation/reader well has three corresponding indicator lights: a red (+), yellow (•) and green
A red light (+), will illuminate and an alarm will sound (if the sound feature is turned on), as soon as a positive RRBI result is detected.
At the end of the specified incubation time, if a negative RRBI result is detected, the green light
After either the green or red light comes on, the RRBI can be removed from the well. The RRBI should not be removed or touched when the yellow light is on.
The positive (unprocessed) control RRBI must provide a positive fluorescent result (red light +). Processed RRBI results are not valid until the positive control reads fluorescent positive (red light +). With the processed RRBI, a positive (red light +) result means a sterilization process failure may have occurred.
The final negative processed RRBI reading for a fluorescence change (green light
If the processed RRBI is further incubated for a visual pH color change, when special studies are desired, the media color change from purple to yellow will also indicate a sterilization failure. The positive control should show a yellow color change at 24 hours (1291) or 48 hours (1292). A final negative color change result (no color change), media remains purple , is made at 24 hours (1291) or 48 hours (1292).
Act on any positive result immediately. The positive result indicates a sterilization process failure may have occurred. Determine the cause of the positive RRBI following current facility policies and procedures. Always retest the sterilizer and do not use the sterilizer for processing loads until 3 consecutive RRBI results are negative.
Disposal
Dispose of the used Attest RRBI according to your healthcare facility policy.
You may wish to sterilize any positive RRBI prior to disposal.
3M © 2003 | Page 6 of 9 | Attest 290 |