INSTRUCTIONS FOR USE
Operating Instructions
Start-up
Place unit on a firm level surface. Plug unit into an appropriate AC grounded outlet.
Allow a 30 minute
It is recommended that the unit be left on to eliminate
Readout Procedure
Follow this procedure for every RRBI to be read:
Step 1. While wearing gloves and safety glasses, CLOSE RRBI cap by pressing down.
Step 2. CRUSH the glass ampule (containing growth media) in the designated crushing well built into the incubator.
Step 3. Hold the RRBI by the cap and TAP bottom of the vial on a tabletop until media wets spore strip at bottom of vial. Do not tap vial on unit.
Step 4. Open the cover and PLACE the RRBI into an incubation/reader well.
Step 5. CLOSE the cover on the Attest 290
WARNING
Do not remove or change placement of RRBI once it is placed into a well. Doing so may result in loss of results and the RRBI test may be invalid.
3M © 2003 | Page 4 of 9 | Attest 290 |