Set the RSB as follows and observe the lamp and audible alarm:

Wrist Strap

29.8Megohms - Red lamp OFF, audible alarm OFF.

40.2Megohms - Red lamp ON (Flashing), audible alarm ON (Chirping).

Ground Clip Disconnect

d)Connect the two single conductor wires to the RSB. Attach the other ends of the wires with clips to the metal jaws of the Model 725 Monitor dual conductor ground clip. Be sure to prevent the clips from contacting each other. Plug in the phone jack test cord into the wrist strap jack on the front of the monitor and short the opposite ends together.

Set the RSB as follows and observe the lamp and audible alarm:

5.0Megohms - Red lamp OFF, audible alarm OFF

11.5Megohms - Red lamp ON (Continuous), audible alarm ON (Continuous tone).

Section 9.

Specifications

Model 725 Monitor Size:

2.5x 2.6 x 1.1 in. (6,4 x 6,6 x 2,8 cm) Accuracy: ± 15%

Test Voltage: 9 Vdc Open circuit

Test Current: Less than 1 microamp

Power Supply Requirements: 9 Volt alkaline battery recommended (Not supplied)

Environmental Operating Conditions:

Temperature: Maximum 110°F, (43°C)

Minimum 50°F, (10°C)

Humidity: Maximum 75% R.H.

Section 10.

Parts Included

1 ea. Model 725 Wrist Strap Monitor with 6 ft. dual conductor ground cord/clip and parking clip.