1-3 Specifications

Table 1 Specifications

Performance specifications

 

 

General

 

 

 

 

 

 

 

 

 

 

 

Power supply

120VAC

60 Hz

(For USA area)

 

 

 

230VAC

50 Hz

(For Europe area)

 

 

 

(factory-preset)

 

 

 

 

 

 

 

 

Power consumption

Approx. 40VA

 

 

 

 

 

 

 

 

Protection against electrical shock

Class I , Type B

 

 

 

 

 

 

 

 

 

Display

LED

 

 

 

 

 

 

 

 

 

EMC

EN 60601-1-2

 

 

 

 

 

 

 

 

 

Blood pressure measurement

 

 

 

 

 

 

 

 

 

Blood pressure measuring method

Oscillometric method

 

 

 

 

 

 

Pressure detection method

Capacitance type pressure transducer

 

 

 

 

 

 

 

Pressure range

0 – 300 mmHg

 

 

 

 

 

 

 

Accuracy

Pressure: ±3 mmHg

 

 

 

Pulse rate: ±5%

 

 

 

 

 

 

 

Measurement range

Blood pressure: 10 – 280 mmHg

 

 

 

Pulse rate: 30 – 200 bpm

 

 

 

 

 

 

 

Pressurizing method

Micro pump

 

 

 

 

 

 

 

Air pressure control method

Rubber valve, ceramic valve

 

 

 

 

 

 

Rapid air exhaust system

Electromagnetic valve

 

 

 

 

 

 

Safety mechanism

The electromagnetic valve is released when

 

 

 

approx. 320 mmHg or greater is detected.

 

 

 

 

 

 

Communications functions

Serial output: RS232C level

 

 

 

 

 

 

Environment specifications

 

 

 

 

 

 

 

 

 

Operating temperature and humidity

10 to 40°C, 85%RH or less, non condensing

 

 

 

 

 

 

Storage temperature and humidity

-20 to 60°C, 95%RH or less, non condensing

 

 

 

 

 

 

Physical specifications

 

 

 

 

 

 

 

 

 

Overall dimensions

245(W) X 320(H) X 390(D) mm

 

 

 

 

 

 

 

Weight

Approx. 9 kg

 

 

 

 

 

 

 

5

Page 13
Image 13
A&D TM-2655P manual Specifications, Performance specifications