7-2 Description of the Functions
Table 5 Functions
No. Function SYSTOLIC
display DIASTOLIC
display Description
F01 f01 off
Not available.
off No printing
1 High-speed printing
2 3-line printing
3 3-line printing
4 Graph printing
F02 f02
5 Table printing
off No bit pattern printing
F03
Printer setting
(See “7-3
Printing
Samples” for
what each
method printing
looks like.)
TM-2655P only f03 on Bit pattern printing
999 The results remain in the display unless
another operation is performed.
5 Displays the results for 5 seconds.
10 Displays the results for 10 seconds.
F04 Displaying time f04
20 Displays the results for 20 seconds.
1 Terminals 1 and 2 to a PC
2 Terminal 1 to a PC
Terminal 2 to a card reader
F05 External
connection Ff05
3 Terminal 1 to a scale
Terminal 2 to a PC
au Automatic pressurization
160 Pressurizes the cuff up to 160 mmHg.
180 Pressurizes the cuff up to 180 mmHg.
F06
Pressurization Ff06
200 Pressurizes the cuff up to 200 mmHg.
120 1200 bps
240 2400 bps
480 4800 bps
F07 Channel 1 baud
rate Ff07
960 9600 bps
120 1200 bps
240 2400 bps
480 4800 bps
F08 Channel 2 baud
rate Ff08
960 9600 bps
F09 Ff09 H9
Not available.
off Does not detect a human presence.
F10 Human sensor Ff10 on Detects a human presence.
F11 Automatic Ff11 off
Does not output the results
automatically.