10. TROUBLESHOOTING
If the
Table 10 Troubleshooting
Problem |
| Check this | Corrective action | ||||
Nothing appears in | Is the power cable connected | Connect the power cable properly. | |||||
properly? | |||||||
the display, even |
| ||||||
|
|
|
|
| Replace both fuses with new ones. | ||
when the power is |
|
|
|
|
| ||
Has a fuse burned out? | (Size 5x20mm. Rated voltage and current | ||||||
turned on. | |||||||
|
|
|
|
| are specified on the rear panel). | ||
|
|
|
|
|
| ||
The cuff will not | Is the arm cuff cover installed | Refer to | |||||
inflate. | with high tension (too tight)? | Cover” to | |||||
| Is your posture proper? | Place your arm at your heart level, relax | |||||
| and remain still during the measurement. | ||||||
|
|
|
|
|
| ||
| Do you remain still? | Do not move your arm during the | |||||
| measurement. Relax and remain still. | ||||||
|
|
|
|
|
| ||
The device will not |
|
|
|
|
| Wearing a thick jacket may cause a faint | |
measure. |
|
|
|
|
| pulse, and result in a measurement error. | |
(Eerr appears.) |
|
|
|
|
| Take off the jacket. | |
|
|
|
|
|
| This device bases its measurements on | |
|
|
|
|
|
| the heartbeat. If you have a very weak or | |
|
|
|
|
|
| irregular heartbeat, the device may have | |
|
|
|
|
|
| difficulty determining your blood pressure. | |
| Is the printer paper installed? | Refer to “6. INSTALLING THE PRINTER | |||||
| (pe appears.) | PAPER” to install the printer paper. | |||||
| Is the printer head up? | Lower the lever and lock the printer paper | |||||
| (HU appears.) | in position | |||||
The printer does | Is the printer paper inserted | Refer to “6. INSTALLING THE PRINTER | |||||
not print. | straight or is the printer paper | ||||||
PAPER” to | |||||||
| jammed? | ||||||
|
| ||||||
|
|
|
|
|
| Refer to | |
| Is the printer setting correct? | Functions” to select the correct printer | |||||
|
|
|
|
|
| setting. |
Note
If the actions described above do not solve the problem, contact the dealer. Do not attempt to repair the device yourself.
23