dealer. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.

Maintenance

Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the A&D service group. The A&D service group will provide technical information, spare parts and units to authorized dealers.

The device was designed and manufactured for a long service life. However it is generally recommended to have the monitor inspected every 2 years, to ensure proper functioning and accuracy. Please contact either your authorized dealer or A&D for maintenance.

Technical Data

Type

UA-767

 

Measurement method

Oscillometric measurement

Measurement range

Pressure:

20 - 280 mmHg

 

Pulse:

40 - 200 beats / minute

Measurement accuracy

Pressure:

±3 mmHg or 2%, whichever is greater

 

Pulse:

±5%

Power supply

4 x 1.5V batteries (R6P or AA)

Upper arm circumference

22 - 32 cm

 

Classification

Type BF

 

Clinical test

According to ANSI / AAMI SP-10 1987

EMC

IEC 60601-1-2: 2001

Operating condition

+10°C to +40°C / 30%RH to 85 %RH

Storage condition

-10°C to +60°C / 30%RH to 85 %RH

Dimensions

Approx. 147 [W] x 64 [H] x 110 [D] mm

Weight

Approx. 300 g, excepting batteries

Note: Specifications are subject to change without prior notice.