and
The HU should be mounted as close as possible to the EBTS to minimize cable losses. Use the following procedure to route and connect the forward and reverse path coaxial cables to the HU:
1.Obtain the required lengths of high performance, flexible, low loss
2.Route the forward and reverse path coaxial cables and the diversity reverse path cable (if the HU supports diversity) between the HU and the EBTS interface (per system design plan) and cut to the required length. Allow sufficient slack for dressing and organizing cables at the HU and for installing an external attenuator in the forward path link.
3.Terminate each cable with an
4.If required, install an external attenuator in the forward path.
Note: The composite forward path RF signal level at the HU must be between
5.Connect the forward path cable to the FORWARD RF IN connector on the HU front panel as shown in Figure 12.
6.Connect the reverse path cable to the REVERSE 1 RF OUT connector on the HU front panel (see Figure 12).
REVERSE 1
RF OUT CONNECTOR
(REVERSE PATH)
REVERSE 2
RF OUT CONNECTOR (DIVERSITY REVERSE PATH)
FORWARD RF IN
CONNECTOR
(FORWARD PATH)
CONNECTOR
Figure 12. Forward and Reverse Path Coaxial Cable Connections
Page 13
© 2004, ADC Telecommunications, Inc.