Appendix A Specifications

A.1 Analog Input

Table A.1: Analog Input

Channels

8 differential

Input type

mV, V, and mA

Input range

J, K, T, E, R, S and B Thermocouple

 

Uni-polar 0~15mV, 0~50mV, 0~100mV,

 

0~500mV, 0~1V, 0~2.5V, 0~20mA,

 

4~20mA

Sampling rate

10 samples/s(total)

Accuracy

±0.1% or better (voltage and current input)

Zero drift

±0.3 uV / °C

 

 

Span drift

±25 ppm / °C

 

 

CMR @ 50/60 Hz

92 dB

 

 

Input impedance

1.8 M

A.2 Accuracy for Thermocouple:

Table A.2: Accuracy for Thermocouple

Input Range

Typical Accuracy

Maximum Error

J thermocouple 0 to 760 °C

±1.0 °C

±1.5 °C

 

 

 

K thermocouple 0 to 1370 °C

±1.0 °C

±1.5 °C

 

 

 

T thermocouple -100 to 400 °C

±1.0 °C

±1.5 °C

E thermocouple 0 to 1000 °C

±1.0 °C

±1.5 °C

R thermocouple 500 to 1750 °C

±1.2 °C

±2.5 °C

 

 

 

S thermocouple 500 to 1750 °C

±1.2 °C

±2.5 °C

 

 

 

B thermocouple 500 to 1800 °C

±2.0 °C

±3.0 °C

 

 

 

NOTE: Due to the location of the CJC sensor, the measurement will have a 1°C max. difference in channels.

USB-4718 User’s Manual

34

Page 40
Image 40
Advantech 4718 user manual Analog Input, Accuracy for Thermocouple