| Table | |||
Current Slew Rate:* |
|
|
| |
Rate # 60 A Range Step | 6 A Range Step | Transition Time | ||
1 | 1 A/ms | 0.1 | A/s | 8.0 ms |
2 | 2.5 A/ms | 0.25 A/s | 3.2 ms | |
3 | 5 A/ms | 0.5 | A/s | 1.6 ms |
4 | 10 A/ms | 1 A/ms | 800 μs | |
5 | 25 A/ms | 2.5 | A/ms | 320 μs |
6 | 50 A/ms | 5 A/ms | 160 μs | |
7 | 0.1 A/μs | 10 A/ms | 80 μs | |
8 | 0.25 A/μs | 25 A/ms | 32 μs | |
9 | 0.5 A/μs | 50 A/ms | 16 μs | |
10 | 1 A/μs | 0.1 | A/μs | 12 μs |
11 | 2.5 A/μs | 0.25 A/μs | 12 μs | |
12 | 5 A/μs | 0.5 | A/μs | 12 μs |
| *AC performance specified from 3 to 60 V. | ||
Voltage Slew Rate: |
|
| |
Rate # Voltage Range Step | Transition Time* | ||
1 | 1 V/ms | 8.0 ms | |
2 | 2.5 V/ms | 3.2 ms | |
3 | 5 V/ms | 1.6 ms | |
4 | 10 V/ms | 800 μs | |
5 | 25 V/ms | 320 μs | |
6 | 50 V/ms | 160 μs | |
7 | 0.1 V/μs | 85 | μs |
8 | 0.25 V/μs | 85 | μS |
9 | 0.5 V/μs | 85 | μS |
*Transition time based on low capacitance current source.
Resistance Slew Rate (1 Ω range): Uses the value programmed for voltage slew rate.
Resistance Slew Rate (1 k and 10 kΩ ranges): Uses the value programmed for current slew rate.
Transient Current Overshoot (When programmed from 0A): |
| ||
Range | Transient Current Level | Current Slew Rate | Overshoot* |
60 A | All slew rates | 0 | |
| 3 A | 1 A/μs to 5 A/μs | 1% |
| 3 A | 1 A/μs to 0.5 A/μs | 0 |
6 A | 6 A | All slew rates | 0 |
| 3 A | 0.25 A/μs to 0.5 A/μs | 1% |
| 3 A | 0.1 A/ms to 0.1 A/μs | 0 |
*Overshoot may be higher during the first five seconds of programming if unit has been operating at full current. Overshoot values assume a total inductance of lμH, or less, in the load leads connected to the D.U.T.
5