Verification

Test Record Form – Agilent N5746A and N5766A

Agilent N5746A and N5766A

Report No _______________

Date __________________

Description

Model

Minimum Specs.

Results

Maximum Specs.

Voltage Programming & Readback

 

 

 

 

Minimum Voltage Vout

Both

20 mV

__________

+ 20 mV

Front Panel Display Readback

Both

Vout 40 mV

__________

Vout + 40 mV

High Voltage Vout

Both

39.96 V

__________

40.04 V

Front Panel Display Readback

Both

Vout 80 mV

__________

Vout + 80 mV

CV Load Effect

Both

6 mV

__________

+ 6 mV

CV Source Effect

Both

6 mV

__________

+ 6 mV

CV Ripple and Noise

 

 

 

 

peak-to-peak

Both

N/A

__________

60 mV

rms

Both

N/A

__________

8 mV

Transient Response

 

 

 

 

Voltage @ 1ms

Both

200 mV

__________

+ 200 mV

Current Programming & Readback

 

 

 

 

Minimum Current Iout

N5746A

19 mA

__________

+ 19 mA

 

N5766A

38 mA

__________

+ 38 mA

Front Panel Display Readback

N5746A

Iout 57 mA

__________

Iout + 57 mA

 

N5766A

Iout 114 mA

__________

Iout + 114 mA

High Current Iout

N5746A

18.962 A

__________

19.038 A

 

N5766A

37.924 A

__________

38.076 A

Front Panel Display Readback

N5746A

Iout 76 mA

__________

Iout + 76 mA

 

N5766A

Iout 152 mA

__________

Iout + 152 mA

CC Load Effect

N5746A

8.8 mA

__________

+ 8.8 mA

 

N5766A

12.6 mA

__________

+ 12.6 mA

CC Source Effect

N5746A

3.9 mA

__________

+ 3.9 mA

 

N5766A

5.8 mA

__________

+ 5.8 mA

 

 

 

 

 

Test Description

N5746A Setting

N5766A Setting

Voltage Programming & Readback, Min Voltage

0

V, 0 A

0

V, 0 A

Voltage Programming & Readback, High Voltage

40 V, 0 A

40 V, 0 A

CV Load Effect, Source Effect, Ripple and Noise

40

V, 19 A

40

V, 38 A

Transient Response

40 V, from 1.9 A to 17.1 A

40 V, from 3.8 A to 34.2 A

Current Programming & Readback, Min Current

0 A, 40 V

0 A, 40 V

Current Programming & Readback, High Current

19

A, 40 V

38

A, 40 V

CC Load Effect, Source Effect

19

A, 40 V

38

A, 40 V

Load Requirements

N5746A

N5766A

 

 

 

Current shunt

0.01100 A

0.01100 A

Agilent N3300 Electronic load modules

2 – N3305A

3 – N3305A

Fixed Resistor for CV Ripple and Noise

2.111kW

1.052kW

 

 

 

104

Series N5700 User’s Guide

Page 104
Image 104
Agilent Technologies N5700 manual Test Record Form Agilent N5746A and N5766A

N5700 specifications

Agilent Technologies, now part of Keysight Technologies, is renowned for its innovative solutions in electronic measurement and instrumentation. Among its impressive portfolio, the Agilent N5700 series of power supplies stands out, providing precision, reliability, and versatility for a range of applications in test and measurement.

The N5700 series features multiple models, each designed to meet the varying power requirements of test systems and electronic devices. One of its key characteristics is its adjustable output voltage and current, allowing users to set parameters according to specific test needs. With outputs ranging from 0 to 60 V and up to 6 A, the N5700 series caters to both low and high-power applications effectively.

A notable technology integrated into the N5700 series is its advanced measurement capabilities. The built-in voltmeter and ammeter enable real-time monitoring of output voltage and current, ensuring precise control over the power supplied to the device under test. This feature is particularly advantageous in troubleshooting and optimization scenarios, providing engineers with immediate feedback on performance.

The N5700 also embraces the latest in power supply control technology with its intuitive user interface. The front panel incorporates a clear display and simple navigation controls, allowing users to program settings easily and access functions without extensive training. Additionally, remote programming capabilities via GPIB, USB, or LAN facilitate integration into automated test setups, enhancing productivity and efficiency.

Thermal management is another hallmark of the N5700 series. Its design ensures effective heat dissipation, enabling reliable operation even under demanding conditions. This robustness is critical in environments where consistent performance is essential, such as in research laboratories and manufacturing facilities.

Furthermore, safety features are thoughtfully included in the N5700 series, such as overvoltage protection, overcurrent protection, and short-circuit protection. These elements assure users that their devices and test setups remain secure during testing, minimizing the risk of damage.

In summary, the Agilent N5700 series power supplies encapsulate the fusion of precision, advanced measurement technology, user-friendly design, thermal management, and robust safety features. This combination makes them an invaluable tool for engineers and researchers engaged in electronic testing and development across various industries. With its versatility and reliability, the N5700 series continues to play a pivotal role in advancing electronic measurement methodologies.