Anritsu MS2026C, MS2038C, MS2028C, MS2036C manual 20 dB

Models: MS2028C MS2036C MS2038C MS2026C

1 24
Download 24 pages 27.3 Kb
Page 4
Image 4

Uncertainty Curves for Round-Trip Cable Loss Measurements (1-Port)

Round-trip cable loss measurements are convenient for field personnel testing installed cable or waveguide runs. This one-port technique provides one-way data after twice traversing the cable. The following two sets of uncertainty curves, less than 6 GHz on the left and greater than 6 GHz on the right, present worst-case uncertainty by DUT Match (i.e., Log Mag) when using VNA Master for one-port cable loss measurements. As a practical tip, consider using a two-port transmission measurement technique to improve upon these one-port cable loss uncertainties.

 

 

 

 

VNA Master (MS20xxC), 5 kHz - 6 GHz

 

 

 

 

 

One Port Cable Loss (Log Mag / 2) Uncertainty vs. DUT Match

 

 

10

 

 

 

 

 

 

 

 

 

(dB)

 

 

 

 

 

 

 

 

 

-20 dB

Uncertainty

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

-25 dB

 

 

 

 

 

 

 

 

 

 

-30 dB

 

0.1

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

 

 

 

 

 

 

Cable Loss, Log Mag / 2 (dB)

 

 

 

 

 

 

 

 

 

VNA Master (MS20xxC), 6 - 20 GHz

 

 

 

 

 

 

 

 

One Port Cable Loss (Log Mag / 2) Uncertainty vs. DUT Match

 

 

 

 

10

 

 

 

 

 

 

 

 

 

 

 

(dB)

 

 

 

 

 

 

 

 

 

 

-20 dB

 

Uncertainty

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

-25 dB

 

 

 

 

 

 

 

 

 

 

 

 

-30 dB

 

 

0.1

 

 

 

 

 

 

 

 

 

 

10

 

0

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

Cable Loss, Log Mag / 2 (dB)

 

 

 

 

These uncertainty curves show how frequency range, DUT Match, and cable loss impact worst-case uncertainty of round-trip cable loss measurements. The uncertainty curves, separated by frequency range, are shown for DUT Match conditions of 20, 25, and 30 dB. For DUT Match of 30 dB and cable loss of 4-5 dB (reflection measurement of 8-10 dB) the worst-case uncertainties are approximately ±1 dB.