Image 8

Low Port Power

3652A Calibration Kit (K-Connector)

Corrected System Performance and Uncertainties: MS202x/3xC Model with 12-term SOLT calibration including isolation using 3652A Calibration Kit

Frequency Range (GHz)

Directivity (dB)

 

 

≤ 5

> 42

 

 

≤ 15

> 36

 

 

≤ 20*

> 32

 

 

* N Connector guaranteed to 18 GHz, typical > 18 GHz

Frequency Range (GHz)

Typical Low Port Power (dBm)

 

 

≤ 3

–25

 

 

≤ 6

–25

 

 

≤ 20

–25

 

 

Measurement Uncertainties

The following graphs provide measurement uncertainty at 23 ºC ± 5 ºC for the above indicated connector type and calibration. Errors are worse-case contributions of residual directivity, source match, frequency response, network analyzer dynamic range, and connector repeatability. For two-port measurements, transmission tracking, crosstalk, and physical load match termination were added. Isolation calibration and an IF Bandwidth of 10 Hz is used.

S11 Magnitude

10

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

1

0.1

 

 

 

 

 

 

 

 

-40

-35

-30

-25

-20

-15

-10

-5

0

S11 Phase

100

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

S21 Magnitude

10

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

1

0.1

0.01

-80

-70

-60

-50

-40

-30

-20

-10

0

S21 Phase

100

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

10

10

1

1

0.1

0.1

 

-40

-35

-30

-25

-20

-15

-10

-5

0

-80

-70

-60

-50

-40

-30

-20

-10

0