Bias Tee (Option 0010)

For tower mounted amplifier tests, the MS20x/3xC series with optional internal bias tees can supply both DC and RF signals on the center conductor of the cable during measurements. For frequency sweeps in excess of 2 MHz, the VNA Master can supply internal voltage control from +12 to +32 V in 0.1 V steps up to 450 mA. To extend battery life, an external power supply can substitute for the internal supply by using the external bias inputs instead. Both test ports can be configured to supply voltage via this integrated bias tees option. Bias can be directed to VNA Port 1 or Port 2.

Frequency Range

 

2

MHz to 6 GHz (MS20x6C)

 

 

2

MHz to 20 GHz (MS20x8C)

 

 

 

 

 

 

 

Internal Voltage/Current

 

+12V to +32V at 450 ma. Steady rate

 

 

 

 

Internal Resolution

 

0.1V

 

 

 

 

 

External Voltage/Current

 

±50 V at 500 mA steady rate

 

 

 

 

 

Bias Tee Selections

 

Internal, External, Off

 

 

 

 

 

 

Port 1

External Bias Input

Port 2

 

 

(±50 V, 500 mA Max)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Internal

 

 

 

 

 

 

 

 

 

Internal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Switch

 

 

 

 

 

 

 

 

 

 

 

 

 

Bias Tee

 

 

 

 

 

Bias Tee

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Internal Bias +12 to +32 V 450 mA Max

RF Test

Switch

Source

 

The VNA Master offers optional integrated bias tee for supplying DC plus RF to the DUT as shown in this simplified block diagram. Connectivity is also provided for external supply (instead of internal) to preserve battery consumption.

Vector Voltmeter (Option 0015)

A phased array system relies on phase matched cables for nominal performance. For this class of application, the VNA Master offers this special software mode to simplify phase matching cables at a single frequency. The similarity between the popular vector voltmeter and this software mode ensures minimal training is required to phase match cables. Operation is as simple as configuring the display for absolute or relative measurements. The easy-to-read large fonts show either reflection or transmission measurements using impedance, magnitude, or VSWR readouts. For instrument landing system (ILS) or VHF Omni-directional Range (VOR) applications, a table view improves operator efficiency when phase matching up to twelve cables. The MS202x/3xC solution is superior because the signal source is included internally, precluding the need for an external signal generator.

CW Frequency Range

5 kHz to 20 GHz

 

 

Measurement Display

CW, Table (Twelve Entries, Plus Reference)

 

 

Measurement Types

Return Loss, Insertion

 

 

Measurement Format

dB/VSWR/Impedance

 

 