( D ) NO DPM INDICATION.
NO DPM INDICATION
Is Display Completely Blank ?
N
Does Display shows only dot Indication ?
Y
Is Continuity between IC7905 &
DPM's OK ?
Y
Is Continuity between ANGD & DPM's OK ?
Y
Is DPM IC OK ?
N
Replace It.
Y | Is Continuity between |
| IC7805 & DPM OK ? |
| N |
| Ensure proper Continuity. |
N | Ensure proper Continuity. |
N | Ensure proper Continuity. |
|
1.If there is no Indication at all ensure that there is proper Continuity between +5V Wire of DPM's and IC7805. It might be possible that +5V Wire Crimping might be loose or IC7805 might be Damaged.
2.It DPM shows only a Dot Indication. Check whether
3.If DPM shows Dot Indication, check whether AGND ( Grey Wire ) is reaching DPM's.
4.If DPM shows Dot Indication, check DPM IC7107. It might be Damaged.