Bayer HealthCare Breeze 2 manual This is a normal expected action

Models: Breeze 2

1 53
Download 53 pages 53.31 Kb
Page 47
Image 47

What You See

What It Means

What You Should Do

 

 

 

 

Meter has detected an

Turn the meter off, then on

 

error; accurate result is

using

, then retest. If the

 

not possible.

error remains, contact Bayer

 

 

 

Customer Service.

or higher

 

 

 

 

 

 

 

 

 

The display

1. You may have

1. Press the

button again.

goes blank when

accidentally pressed

 

 

 

the meter is

the

button.

 

 

 

turned on.

2. Meter has turned

2. This is a normal expected action

 

off after 3 minutes

that saves battery life. Press

 

without use or after

 

to turn on again.

 

10 minutes of no use

 

 

 

 

while waiting for you

 

 

 

 

to apply blood or

 

 

 

 

control solution.

 

 

 

 

 

 

Meter will not

1. Battery life has ended.

1. Replace the battery (see page 35).

turn on.

2. Battery holder is

2. Check the battery and battery

 

 

loose or battery is

holder to be sure they are

 

installed with the

tightly in place and correctly

 

+ side down.

installed (see page 37). If the

 

 

 

display is still blank, replace the

 

 

 

battery (CR2032).

 

 

 

Meter fails to

1. Not enough blood in

1. Discard the used test strip. Retest

start countdown

test strip.

using a round drop of blood.

after blood is

2. There may be debris

2. Inspect the inside of the meter

applied.

inside the meter.

(see page 34) and remove

 

 

 

 

debris if present. Retest.

 

3. Meter failure.

3. Contact Bayer Customer Service.

 

 

 

 

 

 

www.BayerDiabetes.com43